5.0 TESTING PROCEDURES
5.1 Testing Menu Definitions:
T1 Software Displays the software version. Press
PRINT/ENTER
to exit.
Version
T2 Display Flashes all display segments then all indicating Enunciators.
Press
PRINT/ENTER
to exit.
T3 Keys Press the
ZERO
key, b1 will be displayed.
Press the
GRS/NET
key, b2 will be displayed.
Press the
TARE
key, b3 will be displayed.
Press the
lb/kg
key, b4 will be displayed.
Pressing the
PRINT/ENTER
will exit back to the menu
T3
.
T4 A to D Displays Raw counts where a 1mV/V signal from the scale will display
25,000 counts.When the indicator is in angle mode, press the
GRS/NET
key
to cycle Pitch, Roll and Raw counts. Press the
PRINT/ENTER
Key to exit.
T5 Serial Serial Communications can be verified by connecting pins 2 and 3 on the
serial port. A single character will be echoed and pass or fail will be
displayed. Press the
PRINT/ENTER
key to exit.
T6 Setup Data Setup data will be sent out on the printer port.
T7 Default Resets the indicator back to the factory settings. This will clear all calibration
and setup data. “r you sure?” will be displayed. Press the
TARE
key to exit
without Defaulting, or press the
PRINT/ENTER
key to reset the indicator to
the factory default settings.