Ecom EN2 Quick Start Manual - Setup Procedure
SETUP PROCEDURE
1. Connect sample line connections to the analyzer, electrical
connections must be made prior to start-up.
Either
2.
Plug in power supply to ensure analyzer is
fully charged before each use.
3. Power-Up
the
Analyzer.
4. Select Gas Analysis by pressing
5. Now select your fuel type.
Analyzer may prompt “Datalogging?”Press
F4 to choose NO. The EN2 will now do a
self check for 1 minute and then you are
ready to test.
You are now ready to perform an emissions test using the ECOM EN2.
Insert the probe into the designated sampling port, allow the analyzer to draw a sample
for a minimum of 3 minutes so the readings will stabilize.
F1: Stand-by Mode
F2: Quick Print
F3: CO Bypass Pump on/off
F4: Allows user to customize
display values
EXIT to main screen
ENTER to select
POWER KEY
SCROLL Keys
Pressing the UP/
DOWN Keys will
zoom in on the
display
For best results, let the analyzer acclimate to the testing environment for 15 minutes, then
re-start the analyzer. This will allow the analyzer to set a more accurate temperature
compensation and zero reference set point.