Game Ready 550550 User Manual - page 5
EN-4
ModeS oF operAtion
Manual Mode:
The System defaults to this mode, and allows the user to adjust time and
pressure settings.
program Mode:
This mode allows the user to choose one of six programs that provide
treatment for a set time then sleep (no treatment) for a set time, continuously, at a specific
pressure setting.
drain Mode:
This mode allows a user to connect a Hose with a special Hose Adapter to
the unit, enter Drain Mode and have the unit empty the water out of the ice box through
the Hose. Drain Mode can be accessed by pressing the program key and toggling through
all six programs. To empty water in the System while in Drain Mode, press the program
key until you reach Drain Mode, attach the Hose Adapter to the Connector Hose, place the
Hose Adapter over a sink, and press the play button. Drain Mode will run the Control Unit’s
fluid pump for up to six minutes (long enough to fully empty the ice box). Drain
Mode is indicated by the following icons:
d e t A i l e d i n S t r U C t i o n S F o r U S e
BUttonS
power:
Use this button to turn the Control Unit on and off.
program:
Use this button to select one of the available Programs or to return to
Manual Mode. See page 10 in this manual for more information on Programs.
play/pause:
Use this button to start or pause a treatment or to stop.
Add time:
Use this button to add time in Manual Mode (does not work in Program
Mode). You can add up to 90 minutes. Treatment must be paused in order to add or
subtract time.
Subtract time:
Use this button to reduce time in Manual Mode (does not work in
Program Mode). You can subtract up to 90 minutes. Treatment must be paused in
order to add or subtract time.
pressure Selection:
Use this button to select one of four pressure settings: No
Pressure, Low Pressure (5-15 mmHg), Medium Pressure (5-50 mmHg), and High
Pressure (5-75 mmHg). Pressure selection is not available in Program Mode.
Treatment must be paused to change pressure settings.
volume:
Use this button to select the option of sound or no sound. Alarms will still
sound.
C/F Button:
Use this button to select either Celsius or Fahrenheit on the
temperature display.
Backlight:
Use this button to turn the backlight on or off.
Adjusting temperature:
The Control Unit cools the Wrap by pumping water from the ice
box into the Wrap; therefore, the temperature of the Wrap will never be colder than the
water in the ice box. If the knob is turned to three (3) snowflakes, the Control Unit will
provide the coldest possible therapy based on the temperature of the water in the ice box.
Turning the knob towards one (1) snowflake will provide less cold therapy.
note:
The unit should always be started on full cold (3 snowflakes), and run that way for at
least two minutes. The user can then adjust temperature as desired.