4. Startup
4.1 Switching on the device
•
Connect the power supply to the power grid and press the On/Off key (75) on the
release module RM02. The green operational control LED (74) illuminates when
the device is switched on. The camera has no On/Off key and switches on auto-
matically when the PC is switched on.
•
Turn the rotating knob on the slit illumination (77) to a position between '1' and
'10'.
5. Operation
5.1 Setting the eyepieces
NOTE!
Theeyepiecesmustbeindividuallysetpriortothefirstexamination
in accordance with the refraction of the examiner. Insert the adjusting
pin (84) provided in place of the protective cover (85) and turn its black
projection surface at a right angle to the microscope axis. Return the il-
lumination and microscope to the central position (0°).
92. Adjusting pin
93. Protective cover
94. Sliding occluders
94
92
93
•
Each eyepiece should be set individually by turning the knurled ocular refraction
ring with dioptre scale until the projected slit can be seen in focus. The setting is
performedfromthe(+)tothe(-)sideatlowmagnification.
•
The sliding occluders (86) are used to set the correct working distance for the ex-
aminer from the eyepiece.
•
Examiners who do not wear glasses: Pull the occluders out as far as they will go.
•
Examiners who do wear glasses: Push the occluders in as far as they will go.
5.2 Field of view
WARNING!
The images and videos should only be used for documentation purpo-
ses. Only the image in the eyepiece may be used for diagnosis.
Field of view of the object, see table
Circle:
Thefieldofviewoftheobjectobservedthroughthemicroscope's
eyepiece.
Rectangle: Surface area of the sensor:
1 x
1.6 x
ø 18
ø 11.25
21.4 x 18.3
13.4 x 11.4
10 x
6.3 x 5.4
Object image section in the
eyepiece (mm)
Object image section in the
camera with active sensor
area (mm)
5.3 White balance
The image quality is dependent on the correct calibration of the color tones to the
respective slit lamp illumination. We recommend performing a white balance in or-
der to improve the image quality and achieve a realistic color reproduction.
5.3.1 Slit lamp preparation
1. Turn on the slit lamp
2. Filterposition‹open›(nofilter)
3. Setmagnificationto16x
4. Completely open the slit diaphragm
5. Connect the diffuser upstream
6. Position the HAAG-STREIT greycard in front of the slit lamp and use it for fo-
cusing
7. The brightness of the slit lamp's illumination should be set in such a way that
the greycard's structure is clearly discernible.
5.3.2 Conducting a white balance
1. Start the ‹EyeSuite Imaging› software
2. Activate the intensity auto mode
3. Open the ‹White balance› application
4. Start the ‹White balance› by activating the ‹Calibration› function