Pupil distance
54−94mm
Filters
Contrast enhancing (yellow)
Lens
1x
1.6x
1x
1.6x
Eyepiece
10x
10x
25x
25x
Totalmagnification
10x
16x
25x
40x
Objectfielddiameter
18 mm 11.3 mm
8 mm
5 mm
7.4 Instrument base
Operation:
Single-handed operation of control lever in three dimen-
sions
Adjustment of instru-
ment base:
100 mm (length)
100 mm (side)
30 mm (height)
7.5 Dimensions
Weight:
11.5 kg (without power supply, headrest and options)
Dimensions L x W x H: 190 x 320 x 670
Packaging L x W x H: 410 x 490 x 770mm
7.6 Camera
Camera beam
Beam path left (from the point of view of the doctor)
Sensor type
CMOS
Interface
USB 3.0
Frame rate
20 fps (frames per second)
Power consumption
5V / 300 mA
Sensor size
7.2 mm x 5.4 mm
7.7 Minimum PC requirements
Processor type
Intel i5 (or better)
RAM
4 GB (or more with a 64 bit operating system)
Hard disk
At least 500 GB (NTFS data system)
Graphics
Intel HD Graphics 4000 (or better)
Optical drive
DVD-ROM
Monitor
19" with at least 1280 x 800 pixel resolution
Interface
USB 3.0
Operating system
WindowsVista,SP2(32and64bit),Windows7,SP1(32
and 64 bit), Windows 8 / 8.1 (32 and 64 bit)
Sound
Ideally, it is possible to work with sound
8. Maintenance
WARNING!
Do not modify this equipment without authorization of the manufacturer.
Installation and repairs may only be performed by trained specialists.
Contact your HAAG-STREIT representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
The LED illumination can be operated without any maintenance throughout its en-
tire duration of life.
8.1 Device inspection
The following procedure should be used to check that the slit lamp is working cor-
rectly:
•
Insert the test rod in the radial movement bearing, while aligning the surface at a
right-angle to the microscope
•
Set the slit light length to 8 or 14mm
•
Illuminationintensityto50%
•
Setthemagnificationinthemicroscopetomax.
•
Set the eyepieces so that the structure on the test rod is shown in focus. Turn the
eyepiece from the (+) to the (-) side.
•
Forallmagnifications,thestructureofthetestrodmustbeshowninfocus.
•
Close the slit edges to approx. 0.5mm. The edges must be shown in focus.
•
Open the slit edges completely and turn the test rod 45°, the area in focus must
be in the center of the test rod.
8.2 Servicing
To ensure a long service life, the device should be cleaned weekly as described and
covered with a dust cover when not in use. We recommend having the device in-
spected by an authorized service technician annually.