I-SENS e-Checker Owner's Booklet - page 13
22
23
Test results appear after the meter
counts down from five-to-one (5-to-1).
To mark test result, press and hold C
button for three (3) seconds. CHECK
will appear on the display. Marked test
results will not be shown in your
averages.
Compare results with range printed on
test strip bottle. If results fall within
range, you are ready to test your
blood.
Remove used test strip. Meter will turn
off.
Step 3
Step 4
Step 5
Testing Meter and Test Strip Performance
Testing Meter and Test Strip Performance
Insert test strip contact bars facing up
into the meter’s strip port. Gently push
test strip into the port until meter beeps.
A code number will appear for three (3)
seconds.
The
symbol will show on the
display. Confirm matching code numbers
on display and test strip bottle. If code
numbers do not match, see Coding Meter
to Match Test Strips on pages 18-19.
Shake bottle well. Discard first
drop of solution. Wipe off
bottle tip. After
symbol
appears on the display, apply
the solution to narrow top
edge of the test strip until the
meter beeps.
Perform the following steps to check the combined
performance of your e-CheKer Meter and Test Strips or to assess
proper testing procedure performance.
The glucose level printed on the bottle is for the
e-CheKer Control Solution only. It does not refer to your
blood glucose level.
Step 1
Step 2