Icare TONOVET PLUS Instruction Manual - page 6
Icare TONOVET Plus tonometer instruction manual
ENGLISH
Icare TONOVET Plus tonometer instruction manual
ENGLISH
www.tonovet.com
10
www.tonovet.com
11
ACCESSORIES
TECHNICAL INFORMATION
Type: TV011
Dimensions: 24 - 29 mm (W) * 35 - 95 mm (H) * 215 mm (L)
Weight: 140 g (without batteries), 230 g (4 x AA batteries)
Power supply: 4 x AA non-rechargeable batteries, 1.5V alkaline LR6
Measurement range: 10 - 60 mmHg
Accuracy: ± 2.5 mmHg (10 - 30 mmHg) and ± 10% (>30 - 60 mmHg)
Repeatability (coefficient of variation): < 8 %
Accuracy of display: 1 mmHg
Display unit: Millimeter of mercury (mmHg)
The serial number is on the inside of the battery compartment cover.
There are no electrical connections from the tonometer to the patient.
The device has BF-type electric shock protection.
Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30 % to 90 %
Atmospheric pressure: 800 hPa - 1,060 hPa
Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 700 hPa - 1,060 hPa
Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 500 hPa - 1,060 hPa
Mode of operation: continuous
SKU
PRODUCT DESCRIPTION
WEIGHT DIMENSIONS
104
Box of 100 probes
89 g
53 x 109 x 36 mm
7217
Probe base collar
4 g
18 x 18 mm
7218
Probe base collar, narrow
1 g
17 x 18 mm
540
Probe base
4 g
7 x 38 mm
543
Probe base cleaning container
3 g
20 x 56 mm
544
Probe base cover
1 g
19 x 11 mm
559
Wrist strap with lock
4 g
10 x 10 x 270 mm
526
Aluminium case
800 g
240 x 280 x 72 mm
7169
Battery cover & screw
6 g
110 x 25 x 12 mm
565
Silicone grip - green
26 g
45 x 35 x 113 mm
624
IOP pad
38 g
50 x 53 x 16 mm
577
USB memory stick (including instruction manuals as PDF)
44 g
98 x 11 x 93 mm
645
Quick guide
19 g
210 x 148 mm
548
Screw driver
15 g
16 x 90 mm
MEASUREMENT FLOW CHART
Turn Tonometer ON by pressing Select or Measure button
Load Probe
Ready to measure (NOTE the specie setting)
Successful
measurement
with none or
insignificant
deviation.
Tonometer OFF by pressing
Select button > 3 seconds
Measure 6 times by pressing Measure button
(green color bar shows the progress)
This is displayed if you pressed the Select button, pressing of
Measure button would lead you directly to Load Probe
Deviation is too great,
repeat the measurement.
NOTE High IOP may cause high
fluctuation, which may cause
higher standard deviation.
Acceptable measurement
with slightly greater than
normal standard deviation.
Repeating the measurement
is recommended if the IOP is
higher than normal.