GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
Icare TONOVET Plus tonometer (TV011) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Icare TONOVET Plus tonometer (TV011) should assure that it is such an environment.
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment-Guidance
Radiated RF IEC 61000-
4-3
Conducted RF
IEC 61000-4-6
3 V/m
80MHz to 2,5 GHz
3Vrms
150 kHz to 80 MHz
3V/m
NOT APPLICABLE
Portable and mobile RF communications equipment should be used no closer
to any part of the Icare TONOVET Plus tonometer (TV011), including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2 5 GHz
Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
should be less than the compliance level
in each frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the Icare TONOVET Plus tonometer (TV011) is used
exceeds the applicable RF compliance level above, the Icare TONOVET Plus tonometer (TV011) should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the Icare TONOVET Plus tonometer (TV011).
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND
ICARE TONOVET PLUS TONOMETER
Icare TONOVET Plus tonometer (TV011) is intended for use in an electromagnetic environment in which radiated RF-disturbances are controlled. The customer
or the user of the Icare TONOVET Plus tonometer (TV011) can help prevent electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and Icare TONOVET Plus tonometer (TV011) as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of
transmitter
(W)
Separation distance according to frequency of transmitter, m
150 kHz to 80 MHz
d= 1,2 √P
80 MHz to 800 MHz
d= 1,2 √P
800 MHz to 2,5 GHz
d= 2,3 √P
0,01
NOT APPLICABLE
0.12
0.23
0,1
NOT APPLICABLE
0.38
0.73
1
NOT APPLICABLE
1.2
2.3
10
NOT APPLICABLE
3.8
7.2
100
NOT APPLICABLE
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
TV011 is class B equipment and needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided
in Instruction for use manual.
GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC EMISSIONS
Icare TONOVET Plus tonometer (TV011) is intended for use in the electromagnetic environment specified below.
The user of the Icare TONOVET Plus tonometer (TV011) should assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
Icare TONOVET Plus tonometer (TV011) is battery operated and use RF energy only for its internal
function. Therefore, its RF emissions are low and are not likely to cause any interference in nearby
equipment.
RF emissions CISPR 11
Class B
Icare TONOVET Plus tonometer (TV011) is suitable for use in all establishments, including domestic
establishments and those directly connected to public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Voltage fluctuations flickering
emissions
IEC 61000-3-3
NOT APPLICABLE
GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC IMMUNITY
Icare TONOVET Plus tonometer (TV011) is intended for use in the electromagnetic enviroment specified below.
The customers or users of Icare TONOVET Plus tonometer (TV011) should assurre that it is used in such enviroment.
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment-Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
±8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %
Electrical fast Transients/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
NOT APPLICABLE
NOT APPLICABLE
Surge
IEC 61000-4-5
±1 kV for line(s) to line(s)
±2 kV for line(s) to earth
NOT APPLICABLE
NOT APPLICABLE
Voltage dips, short interruption
and voltage variations on power
supply lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
NOT APPLICABLE
NOT APPLICABLE
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.