Icare TONOVET PLUS Instruction Manual - page 7
Icare TONOVET Plus tonometer instruction manual
ENGLISH
Icare TONOVET Plus tonometer instruction manual
ENGLISH
www.tonovet.com
12
www.tonovet.com
13
SYMBOLS
ELECTROMAGNETIC DECLARATION
WARNING
Use of any accessories and cables other than those specified in the manufacturer’s documentation, with the exception of cables sold by the manufacturer as
replacement parts for internal components, may result in increased emissions or decreased immunity of the Icare TONOVET Plus tonometer.
WARNING
Use of any accessory or cable with the Icare TONOVET Plus tonometer other than those specified may result in increased emissions or decreased immunity of the
Icare TONOVET Plus tonometer.
SN
LOT
Attention!!! See instructions
See operating instructions
for more information
Serial number
Single use only
BF-type device
Lot number
Manufacturing date
Keep dry
Manufacturer
Storage environment
Transport environment
PERIODIC SAFETY CHECKS
We recommend that the following checks be performed every 24 months.
Equipment inspection for mechanical and functional damage.
Inspection of safety labels for legibility.
MAINTENANCE
Follow local regulations and recycling instructions regarding the disposal or recycling of the Icare TONOVET tonometer and accessories.
WARNING
The tonometer should only be opened by qualified service personnel. It contains no userserviceable parts, apart from the batteries and a probe base.
The Icare TONOVET tonometer requires no routine servicing or calibration other than changing the batteries at least every 12 months and changing or cleaning
the probe base. If servicing is necessary, contact your local Icare representative.
REPLACING/CLEANING THE PROBE BASE
During the use some dirt may collect in the probe base, affecting the probe movement. Replace the probe base every 12 months. Clean the probe base
every 6 months. Clean or replace the probe base if the error message Clean Change is displayed.
Instructions for replacing the probe base:
• Replace every 12 months.
• Turn off the tonometer.
• Unscrew the probe base collar.
• Remove the probe base by tilting the tonometer downwards
and pulling the probe base out of the tonometer.
• Insert a new probe base into the tonometer.
• Screw the collar in, to lock the probe base.
Instructions for cleaning the probe base
:
• Clean every 6 months.
• Fill the probe base cleaning container or other clean container with 70-100 % isopropyl alcohol.
• Turn the power off.
• Unscrew the probe base collar.
• Remove the probe base by tilting the tonometer downwards
and pulling the probe base out of the tonometer.
• Insert the probe base into the cleaning container and let soak for 5–30 minutes.
• Remove the probe base from alcohol.
• Dry the probe base by blowing clean canned or compressed air into the hole
in the probe base. This will additionally remove possible residual dirt.
• Insert the probe base into the tonometer.
• Screw the collar in, to lock the probe base.
CLEANING THE TONOMETER
WARNING
Never immerse the Icare TONOVET tonometer, spray, pour or spill liquid onto the Icare TONOVET tonometer, its accessories, connectors, switches or open-
ings in the chassis. Dry any liquid on the surface of the tonometer immediately.
Icare TONOVET tonometer’s surfaces have been tested and found chemically resistant to the following liquids:
• 70-100 % isopropyl alcohol
• Mild soap solution
• 95 % Pursept solution
Cleaning instructions for surfaces:
• Turn the power off.
• Dampen a soft cloth with one of the liquids mentioned above.
• Lightly wipe the surfaces of the tonometer with the soft cloth.
• Dry the surfaces with a dry soft cloth.
RETURNING THE ICARE TONOVET TONOMETER FOR SERVICING / REPAIR
Contact your local Icare representative for shipping instructions. Unless otherwise instructed by Icare Finland, there is no need to ship accessories along
with the tonometer. Use a suitable carton with the appropriate packaging material to protect the device during shipment. Return the device using any
shipping method that includes proof of delivery.
Do not discard this product with other household-type waste. Send to appropriate
facility for recovery and recycling. EU WEEE (European Union Directive for Waste of
Electronic and Electrical Equipment)
Temperature limits
Humidity limits
Atmospheric pressure limits
-10°C
55°C
95%
10%
1060hPa
700hPa
-40°C
70°C
95%
10%
1060hPa
500hPa