Jadever JWI 700C Instruction Manual - page 11
JWI-700C
10
6-1 Linear calibration
1)Turn on the power while pressing key
TARE
to initiate linear calibration.
2) Again press key
TARE
to enter zero calibration
mode, with “
”flashing on the weight window.
3) Wait till “
” appears put weights of 1/3
of full load on the weighing pan and press down key
TARE
. (E.g. for 75kg weighing capacity, 1/3 of full
load is 25kg)
4) Wait till
appears, then put weights
of 2/3 of full load(50kg)on and press down
key
TARE
.
5) Wait till “
” appears, then put weights
of full load(75kg) on and press key
TARE
.
6) The calibration Procedure is completed with a
symbol of “
”flashing, then take away the
weights.
7)Press key
TARE
to return to weighing mode.
6-2 Single point calibration
1) Turn on the power while pressing key
SMPL
to
initiate single point calibration.
2) Input 11 with numeric keys
3) Again press key
SMPL
to enter the zero point
calibration mode, with “ ”flashing.