Kaltis ULTRA LOW TEMPERATURE FREEZER User Manual - page 10
WARNING!
Do not use outdoors. Current leakage or electric shock may result if
the unit is exposed to water.
Never install the unit in an excessively humid environment. or where
there is a high risk of being splashed. by water. Deterioration of the
insulation may result which could cause current leakage or electric
shock.
CAUTION!
To allow for proper air flow, a minimum of fifteen centimeters clearance is
required behind the freezer.
5.2 Wiring
CAUTION!
Connect the equipment to the correct power source. Incorrect voltage can
result in damage to the equipment.
CAUTION!
For personal safety and trouble-free operation, the unit must be properly
grounded. Failure to ground the equipment may cause personal injury or
damage to the equipment. If the power supply outlet is not grounded, it is
necessary to install a ground from the ground screw located on the lower
right rear side of the unit by qualified engineers.
If possible, it is preferable to connect the freezer to a dedicated (separate) power
supply.. Each freezer is equipped with a service cord and plug designed to connect it
to a power outlet which it delivers the correct voltage. Supply voltage must within
+10% to -10%
of the freezer’s rated voltage.
CAUTION!
Never cut or remove the grounding prong from the service cord plug. If the
prong is removed, the freezer will not be properly grounded.
10