Keithley 707B Quick Start Manual - page 16
Test the instrument
To test the instrument for basic operation:
1.
On the front cover of the instrument, press the
SLOT
key. You
should see the word "MAIN:" and the model number,
description, and the firmware revision for the instrument.
2. Turn the navigation wheel to display the firmware version
information for the installed switching card(s).
3. Press
the
EXIT
key to return to the operating display.
4. To close a channel, push the navigation wheel to edit the slot and
row, turn the wheel to change the value. Push the navigation
wheel to accept the displayed slot and row; this will also activate
edit mode for the column. Rotate the navigation wheel to
change the column value and push the navigation wheel to accept
the channel shown on the display.
5. Press
CLOSE
to close the channel. The closure should be
indicated on the display.