Kusam-meco KM 061 Operation Manual

Manual is about: digital clampmeter and multimeter

Summary of KM 061

  • Page 1

    17, bharat industrial estate, t. J. Road, sewree (w), mumbai-400015. India sales direct: (022)24156638 tel.:(022) 2412 4540, 2418 1649 fax: 2414 9659 e-mail : kusam_meco@vsnl.Net, website : www.Kusamelectrical.Com; www.Kusam-meco.Co.In list of products * digital multimeter * digital ac & ac/dc clamp...

  • Page 2

    Nearly every electrical engineer has a hand held digital clamp meter (tongtester). We sometimes take them for granted, until we damage them or “burn them out”. If you incorrectly connect your clamp meter to a circuit, or if you have the clamp meter on wrong setting, you damage the meter and possibly...

  • Page 3

    -1- safety this manual contains information & warnings that must be followed for operating the instrument safely and maintaining the instrument in a safe operating condition. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be imp...

  • Page 4

    7 7 -3- 2006/95/ec and electromagnetic compatibility directive 2004/108/ec -4- average sensing (km 061) true rms sensing (km 062) safety : meets iec61010-1 2nd ed., en61010-1 2nd ed., ul61010-1 2nd ed., can/csa c22.2 no. 61010.1-0.92, iec61010-2-032, en61010-2-032 & ul61010b-2-032 & can/csa c22.2 no...

  • Page 5

    -5- e.M.C. : meets en61326-1:2006 (en55022, en61000-3-2, en61000-3-3, en61000-4-2, en61000-4-3, en61000-4-4, en61000-4-5, en61000-4-6, en61000-4-8, en61000-4-11) in anrf field of 3v/m : capacitance function is not specified other function ranges : total accuracy = specified accuracy + 50digits perfo...

  • Page 6

    -7- resistance (ohms) range resolution accuracy 400.0 w 4.000 kw 40.00 kw 400.0 kw 4.000 mw ±(0.8%rdg + 6dgts) ±(0.6%rdg + 4dgts) ±(1.0%rdg + 4dgts) open circuit voltage : 0.4v dc typical 40.00 mw ±(2.0%rdg + 4dgts) audible continuity tester open circuit voltage : 0.4v dc typical range : 400.0w ; ac...

  • Page 7

    -9- hz frequency function display counts : 5000 maximum resolution : 0.001hz accuracy : 0.5% + 4d 400.0 mv 4.000 v 40.00 v 400.0 v sensitivity (sine wave) range 5hz ~ 20khz 5hz ~ 100khz 10hz ~ 1khz 5hz ~ 100khz 5hz ~ 5khz 600 v 350 mv 3.2 v 25 v 100 v 410 v dca / aca unspecified -10- 3) product desc...

  • Page 8

    -12- note : dc 400.0mv range is designed with 1000mw high input impedance for least current drain in measuring small signals, and can cope better with most commercially available voltage output transducers / adapters. The non-zero display reading is normal when the meter inputs are open circuit, whi...

  • Page 9

    -13- caution (application and removal of the clamp-on meter) press the jaw trigger and clamp the jaws around only one single conductor of a circuit for load current measurement. Make sure the jaws are completely closed, or else it will introduce measurement errors. Enclosing more than one conductor ...

  • Page 10

    -15- on on select inputs are made through the test leads terminals. Slide-switch on defaults at w . Press select button momentarily 3 times to select capacitance function. Relative zero mode can be used to zero out the parasitic capacitance of the leads and the internal protection circuitry of the m...

  • Page 11

    -17- -16- -18- the meter uses standard 1.5v aaa size battery x 2; or 1.5v aaa size alkaline battery x 2. Loosen the 2 captive screws from the battery cover case. Lift the battery cover case. Replace the batteries. Replace battery cover case. Re-fasten the screws..

  • Page 12: Test Certificate

    Iso 9001 registered qc pass kusam-meco mumbai test certificate digital clampmeter + multimeter this test certificate warrantees that the product has been inspected and tested in accordance with the published specifications. The instrument has been calibrated by using equipment which has already been...