Changing the
Silhouette® Infusion Set Using a Paradigm® REAL-Time Insulin Pump
(models 522, 722)
1
To Fill the Reservoir:
To Rewind the Piston in the Pump:
1. Wash your hands. Remove the old infusion set
from your body and the reservoir from the pump.
2. Select Prime in the
MAIN MENU.
Then select Rewind
to open the
REWIND screen.
Press
ACT
to rewind.
1. Remove the
reservoir from the
package.
2.Pull the plunger down
to fill the reservoir with
air.
WARNING:
If
insulin, or any
liquid, gets
inside the tubing
connector, it can
temporarily block
the vents that
allow the pump
to properly prime the infusion set. This may
result in the delivery of too little or too much
insulin, which can cause hypoglycemia or
hyperglycemia. To prevent insulin from getting
inside the Silhouette tubing connector, after you
fill the reservoir make sure you hold the insulin
vial upright when you remove the reservoir from
the transfer guard. If you do not hold the insulin
vial upright, insulin can get on the top of the
reservoir and could transfer liquid into the tub-
ing connector.
4. Push down on the
plunger to push air
from the reservoir
into the vial.
Continue to
hold the plunger
down.
5. Flip the vial over
so it is now on top.
Slowly pull the
plunger down to
fill the reservoir.
6. Check the
reservoir for air
bubbles. Tap the
side of the reservoir
to force any bubbles
to rise to the top.
Push the plunger
up to move the air bubbles from the reservoir
into the insulin vial.
7. After the air bubbles
are removed, slowly
pull the plunger
down to fill the
reservoir with enough
insulin to last 2-3 days.
8. To avoid getting insulin
on the top of the
reservoir, flip the vial
over, so the vial is
upright. Hold the
transfer guard, and turn
the reservoir counter-
clockwise. Pull the
reservoir straight up to
disconnect it from the transfer guard. Discard
the transfer guard into a sharps container.
3. Wipe the top of the
insulin vial with
alcohol. Hold the
blue transfer guard
and press it
down onto
the insulin vial.
Tubing connector
Top of
reservoir
plunger
rod
transfer
guard
reservoir
vial
3
1