Changing the
Silhouette® Infusion Set Using a Paradigm® REAL-Time Insulin Pump
(models 522, 722)
To Insert the Silhouette® Using the Sil-Serter®:
6. Rest the arms of the
Sil-serter against your skin
so that the serter is at a
20-45 degree angle.
7. Press the white button at
the top of the serter to
insert the infusion set.
8. Hold the Silhouette
against your skin with
one hand. With the other
hand, press the Lock &
Release handle to open
the serter. Then, slide the
serter back from the
Silhouette.
4
Par
adig
m, Silhouett
e and Sil-ser
ter ar
e r
eg
ist
er
ed tr
ademar
ks of M
edtr
onic M
iniM
ed
, I
nc
. 950M11373-011 20130610 ©M
edtr
onic M
iniM
ed
, I
nc
. 2013. All r
igh
ts r
eser
ved
.
To Remove the Introducer Needle:
1. Peel the paper from the
front part of the adhesive
and press the adhesive
securely against your skin.
2. Squeeze the arms on the
gray needle handle and
pull it back to remove the
introducer needle.
3.Peel the rest of the paper
without lifting the
infusion set too far from
your skin. Press the
adhesive securely to
your skin.
4. Connect the infusion set
tubing by sliding the site
connector into the
insertion piece on your skin.
1. From the PRIME MENU select Fixed Prime and
press
ACT
.
2. If you are using a
6mm cannula,
press to enter
0.7 units of insulin.
Press
ACT
.
To Prime the Cannula: