OHAUS CHAMP II BASE Operating And Installation Instructions

Other manuals for CHAMP II BASE: Operating And Installation Instructions
Manual is about: with CD11

Summary of CHAMP II BASE

  • Page 1

    Ohaus corporation 29 hanover road, p.O. Box 900 florham park, nj 07932-0900 phone: 973-377-9000 fax: 973-593-0359 champ ii scale with cd11 operating and installation instructions.

  • Page 2

    2 ohaus corporation, 29 hanover road, florham park, new jersey, 07932, usa declaration of conformity we, ohaus corporation, declare under our sole responsibility that the instrument models listed below marked with “ce” - are in conformity with the directives and standards mentioned. Konformitätserkä...

  • Page 3

    3 note: this equipment has been tested and found to comply with the limits. For a class a digital device, pursuant to part 15 of the fcc rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equ...

  • Page 4

    4 champ ii bench scale 1. Overview champ ii is a single load cell bench scale connected to cd-11 industrial indicator. 2. Physical descriptions construction platform - fabricated stainless steel scale base - formed and welded mild steel, painted black. Overloading corner loading - 100% of full capac...

  • Page 5

    5 3. Installation procedures assembly and installation examine the shipping box for any signs of damage. If damage is found, make a claim with the carrier immediately. Open the box and remove the scale assembly. Place the scale assembly on a stable flat surface. Follow the “champ ii scale assembly i...

  • Page 6

    6 load cell replacement remove the scale platform and disconnect the indicator battery and ac power source. Important note wait 30 seconds after removing power to the indicator before unplugging the load cell cable. Remove the top load cell mounting bolts that secure the top frame to the load cell. ...

  • Page 7

    7 figure 1-1 figure 1-2 position ch15r ch30r ch60r ch60l ch100r ch150r ch300r a 2mm (0.078”) 2mm (0.078”) 4mm (0.157”) 4mm (0.157”) 4 mm (0.157”) b 0.5mm (0.02”) 0.5mm (0.02”) 0.75mm (0.03”) 0.75mm (0.03”) 1mm (0.04”) table no. 1 overload gap settings shift test a shift test verifies that all sectio...

  • Page 8

    8 positions a, b, c, and d are centered at the each quarter of the scale platform. The following table shows the tolerance in “d” (division) for the shift test. Scale capacity test weight acceptable tolerance (new scale) maintenance tolerance (in service) 3,000d 1,000d "1.0d "2.0d table no. 2 tolera...

  • Page 9

    9.

  • Page 10

    10 champ ii bench scale specifications model ch15r11 ch30r11 ch60r11 ch60l11 ch100r11 ch150r11 ch300r11 capacity x readibility (kg)* 15 x 0.002 30 x 0.005 60 x 0.01 60 x 0.01 100 x 0.02 150 x 0.02 300 x 0.05 (lb) 30 x 0.005 60 x 0.01 120 x 0.02 120 x 0.02 200 x 0.05 300 x 0.05 600 x 0.1 factory set ...

  • Page 11

    11 warranty ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided th...