OHAUS CHAMP II BASE Operating And Installation Instructions - page 5
5
3. Installation Procedures
Assembly and Installation
Examine the shipping box for any signs of damage. If
DAMAGE is found, make a claim with the carrier
immediately.
Open the box and remove the scale assembly. Place
the scale assembly on a stable flat surface.
Follow the Champ II Scale Assembly Instructions.
Level the scale, apply power and weigh.
Good
Wrong
Level Indicator
BEFORE CALIBRATION OR USE, REMOVE THE
RED PLASTIC SHIPPING TABS IN THE CORNERS
OF THE SCALE BASE.
Troubleshooting
If operational difficulties are encountered, first obtain
as much information as possible regarding the
problem. Failures and malfunctions often may be
traced to simple causes such as loose connections to
the indicator, low battery power, improper setup, etc.
If simple causes cannot be found, additional
troubleshooting is best performed by substitution. A
PCB or LOAD CELL believed to be defective may be
checked by replacing the suspect part with known
good part and then observing whether the problem is
corrected.