OHAUS CHAMP II BASE Operating And Installation Instructions - page 7
7
Figure 1-1 Figure 1-2
Position
CH15R
CH30R
CH60R
CH60L
CH100R
CH150R CH300R
A
2mm
(0.078)
2mm
(0.078)
4mm
(0.157)
4mm
(0.157)
4 mm
(0.157)
B
0.5mm
(0.02)
0.5mm
(0.02)
0.75mm
(0.03)
0.75mm
(0.03)
1mm
(0.04)
Table No. 1 Overload Gap Settings
Shift Test
A shift test verifies that all sections of the scale platform weigh
within tolerance. If the scale does not pass the shift test, verify
the overload stop gaps before replacing the load cell. No
adjustment for the shift is possible. If the shift test cannot be
passed, the load cell must be replaced.
Place test weights equal to one third scale capacity
sequentially at each of the positions A, B, C, D, as shown in
Figure 1-3. Note the indicator reading at each position.
A
C
D
B
Figure 1-3 Shift Test
A
A
A
A
B
B
Overload Stop
Jam Nut
Set Screw
Gap