Quell QE-001 User Manual - page 37
Quell
®
User Manual Quell Customer Care 1-800-204-6577 or +1-781-890-9989
<
Contents |
36
APPENDIX I: Electromagnetic Compatibility Declaration
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions; CISPR 11
Group 1
Quell uses RF energy only for its internal function. Its
RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions; CISPR 11
Class B
Quell is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonics Current
Emissions IEC 61000-3-2
Class A
Fluctuations and Flicker
IEC 61000-3-3
Complies
Electrostatic Discharge
Immunity (ESD);
IEC 61000-4-2
+/- 15kV air; +/- 8kV contact
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
humidity should be at least 30%
Radiated RF;
IEC 61000-4-3
10 V/m; 80 MHz to 2.7 GHz
10 V/m compliance level
Electric Fast Transient
Burst Immunity
IEC 61000-4-4
±2kV power leads
Mains power quality should be that of a typical
domestic establishment.
Fast Surge Immunity
IEC 61000-4-5
±2kV Common mode
Radio Frequency
Common Mode Immunity
IEC 61000-4-6
3 VRMS (80% AM at 1 kHz);
150 kHz to 80 MHz
Power Frequency
Magnetic Field Immunity
IEC 61000-4-8
30 A/m 50 Hz
Voltage Dip and Interrupt
Immunity IEC 61000-4-11
Complies
Mains power quality should be that of a typical
domestic establishment.
APPENDIX I
Electromagnetic Compatibility Declaration
Quell is intended for use in the electromagnetic environment specified below. The user
should ensure that it is used in such an environment.