Radiancy Kyroback User Manual - page 17
15
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity – Kyrobak
The Kyrobak is intended for use in the electromagnetic environment specified below; The customer or the user of the Kyrobak should assure that it is used in such an environment.
Immunity test
IeC 60601-1-2
Test level
Compliance level
electromagnetic environment - guidance
electrostatic discharge
(eSD)
IeC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or Ceramic tile. If floors are covered with synthetic material,
the relative humidity Should be at least 30 %.
electrical fast transient/burst
IeC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
not Applicable
Mains power quality should be that of a typical public low-voltage power supply network that
supplies buildings used for domestic purposes, commercial or hospital environment.
Surge
IeC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical public low-voltage power supply network that
supplies buildings used for domestic purposes, commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IeC 61000-4-11
<5 %UT (>95 %dip in UT)
for 0,5 cycle
40 %UT (60 %dip in UT)
for 5 cycles <5 %UT
70 %UT (30 %dip in UT)
for 25 cycles <5 %UT
<5 %UT (>95 %dip in UT)
for 5 s
<5 %UT (>95 %dip in UT)
for 0,5 cycle
40 %UT (60 %dip in UT)
for 5 cycles <5 %UT
70 %UT (30 %dip in UT)
for 25 cycles <5 %UT
<5 %UT (>95 %dip in UT)
for 5 s
Mains power quality should be that of a typical public low-voltage power supply network that
supplies buildings used for domestic purposes, commercial or hospital environment.
If the user of the Kyrobak requires continued operation during power mains interruptions; it is
recommended that Kyrobak be powered from an uninterruptible power Supply or a battery.
Power frequency (50/60
Hz)magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical public
low-voltage power supply network that supplies buildings used for domestic purposes,
commercial or hospital environment.
nOTe: UT is the a.c. mains voltage prior to application of the test level.