Rae QRAE 3 User Manual - 9. Policy Enforcement
30
9. Policy Enforcement
The QRAE 3 can be configured to enforce a facility/company’s requirements that
calibration and/or bump testing be performed at specified intervals, and to warn the user
that calibration/bump testing is required. In addition, it can require calibration or bump
testing and not allow normal operation of the instrument unless calibration or bump
testing is performed.
If Policy Enforcement is enabled and if the instrument has been bump tested and
calibrated in compliance with the policy settings, a check-mark icon is included along the
top of the QRAE screen:
If Policy Enforcement is enabled, then after startup the QRAE 3 displays a screen that
informs the user that the instrument requires either a bump test or a calibration. If both
are required, then they are shown in sequence.
9.1. Setting Policy Enforcement
You must use ProRAE Studio II to make changes to Policy Enforcement settings.
Make sure the AC adapter is connected and that a USB cable is connected to the Travel
Charger and a computer running ProRAE Studio II.
1. Turn on the QRAE 3. Allow the system to start up and go through its startup routine.
2. Press [MODE] until “Enter Communications Mode?” is displayed.
3. Press [Y/+]. The screen now displays: “Ready To Communicate With Computer.”
4. Start ProRAE Studio II.
5. Select “Administrator.”
6. Input the password (the default is “rae”).
7. Click “OK.”
8. Click “A” (detects instruments automatically).
9. Click on the instrument’s icon when it appears to highlight it.
10. Click “Select.”
11. Click “Setup.”
12. Click “Policy Enforcement.”