Sensidyne Gilian 3500 Operation Manual - Specifications
23
REF 360-0039-01 (B)
APPENDIX B
SPECIFICATIONS
Performance
Operating High Flow Range ........................ 700–3500 cc/min
Accuracy (Air Flow Readout) .......................
±
5%
Constant Flow control ..................................
±
3% of set flow (after calibration)
between 1-3 LPM;
±
5% of set flow below 1 LPM
and over 3 LPM
Constant Flow Compensation ...................... 3500cc up to 12” water back pressure
3000cc up to 20” water back pressure
2500cc up to 25” water back pressure
2000cc up to 30” water back pressure
1000cc up to 30” water back pressure
700cc up to 40” water back pressure
Run Time .......................................................
8 hour minimum (See Chart)
Flow Fault ..................................................... If flow changes exceed 5%, fault icon appears.
If fault exceeds 30 seconds, pump shuts down.
Pump attempts to restart every 3 minutes for
up to 1 hour.
General
Controls ......................................................... Power/Enter, Set/Cal,
▲
/Clear,
▼
/Run/Stop
Indicators ...................................................... Green LED (“Normal Operation”)
Icons (LCD) ................................................... Battery Indicator, Hold, Fault, Set
Dimensions ................................................... 3.2” (W) x 5.4” (H) x 2.3” (D)
Weight ........................................................... 19.5 oz.
Display (Normal Operation) ........................ Live Flow, Elapsed Time & Volume Sampled
Electrical
Battery Pack .................................................. Removable, Sealed,
Rechargeable Nickel-Metal-Hydride (4.8V, 2.2 AH)
Battery Level Indicator ................................. Icon displays Full, Mid, & Low charge levels
Interface Connectors .................................... Charging Jack
Charge Time ..................................................
Approvals
Intrinsic Safety
UL Certification ............................................. Class I, Div 1, Groups A, B, C, D;
Class II, Groups E, F, & G; Class III. T3C.
C
UL Certification ........................................... Class I, Div 1, Groups C, D;
Class II, Groups E, F, & G; Class III. T3C.
CE
CENELEC Certification .................................. ATEX Ex II 2 G, EEx ib IIB T4
EMC EMI/RFI ................................................ EN 55022 Class B; EN 50082-1; IEC 801-2, 3, & 4
Environmental
Temperature
Operating ...................................................... 0
°
C to 45
°
C (32
°
F to 113
°
F)
Storage ........................................................... -20
°
C to 45
°
C (-4
°
F to 113
°
F)
Charging ........................................................ 5
°
C to 40
°
C (41
°
F to 104
°
F)
Humidity
Operating ...................................................... 0–85 %RH
Storage ........................................................... 0–98 %RH, non-condensing