Alerts, Alarms, and Troubleshooting Table
Sh
U
tdo
W
n
cond
It
Ion
S
d
Ev
Ic
E
n
EE
d
S
a
tt
Ent
Ion
typ
Ical
cond
It
Ion
S
gREEn
IndIcatoR
yElloW
IndIcatoR
REd
IndIcatoR
aUdIblE
alaRm
What to do
condItIon
“Warming Up. Please
Wait.”
On
On
On
Off
Wait. The system typically takes 3 - 5
minutes to reach specified performance.
“3.0” (or flow setting)
On
Off
Off
None
Nothing. The system is operating properly
at the specified flow rate.
“Low or Warm Power
Cartridge”
On
Blinking
Off
Beeps
Plug into external power or replace with a
charged Power Cartridge.
“9-Volt Battery”
On
On
On
3 Beeps
The internal 9-volt battery needs replaced.
Contact dealer/provider for replacement.
“O
2
Concentration
Off
Blinking
Off
Off
Continue Use. Check or clean air intake
filter. If condition persists, contact
dealer/provider.
“O
2
Concentration
Off
Off
Blinking
Beeps
Continue Use. Check or clean air intake
filter. If condition persists, contact
dealer/provider.
“Flow Rate Error
/Blocked Flow”
Off
Blinking
Off
Beeps
Check tubing and/or humidifier for
obstruction. Check/clean air intake filter. If
condition persists past 10 minutes, contact
dealer/provider.
“Cannot Charge Power
Cartridge”
Off
On
Off
Off
Seek external power source. Remove Power
Cartridge and allow to cool to room tem-
perature. If Power Cartridge malfunction
persists, contact home care provider.
“No Inspiratory Effort
Detected” (in Pulse
Flow Mode only)
On
On
Off
On
Make sure tubing does not exceed 7 ft
(2.1m) when using Pulse Flow Mode.
Audible alarm will beep after 45 seconds
if no inspiratory effort is detected. If no
inspiratory effort is detected 15 seconds
after that, the Eclipse will automatically
switch to Continuous Flow Mode
of operation.
“Power Cartridge
Shutdown or hot
Power Cartridge”
(with no external
power present)
Off
Off
Blinking
On
Seek external power source. Remove Power
Cartridge and allow to cool to room tem-
perature. If Power Cartridge malfunction
persists, contact home care provider.
“Loss of External
Power” (without
charged Power
Cartridge installed)
*Note: Stars will be
flashing in the LCD
********.”
Off
Off
Blinking
On
Seek an alternative oxygen supply until
external power is restored or install a
charged Power Cartridge.
“System Malfunction”
Off
Off
On
On
10 seconds
then silent
Contact home care provider. Use back up
oxygen supply.