Personal Ambulatory Oxygen System
Users Manual
Eclipse
Optional Accessories
Check with your home care provider,or visit us at www.sequal.com for more information about
optional accessories. There are many different types of oxygen tubing, cannulae, and humidi-
fiers. The following items are recommended by SeQual Technologies for use with
your Eclipse.
Salter labs
®
humidifier, part number 7600, or equivalent: If your physician has pre-
scribed an optional humidifier, follow the manufacturer’s instructions for use. Attach the
humidifier to the oxygen outlet port of the Eclipse. Use of optional humidifiers not recom-
mended for the Eclipse may impair performance of the device and may void the warranty.
do not use a humidifier in the Pulse Flow Mode. The Eclipse will not detect inspiratory effort.
The device will alarm and default to the Continuous Flow Mode for continuing operation.
SeQual technologies humidifier adapter – part number 3509: If your physician has
prescribed an optional humidifier, you may need to use the SeQual Humidifier Adapter.
Follow the instructions for use. Attach the Humidifier Adapter to the oxygen outlet port
of the Eclipse and then to the humidifier. Attach the cannula, or oxygen tubing to the
humidifier outlet.
Salter labs oxygen Supply tubing, part number Series 2000, or equivalent: The inter-
nal diameter should be no less than 3/16” (0.48 cm). Connect the oxygen tubing to the outlet
port of the humidifier, or directly to the oxygen outlet port of the Eclipse if you do not use a
humidifier. Connect the other end of the tube to the nasal cannula, if oxygen supply tubing is
not already attached to the cannula. Tubing not specified for use with this Eclipse may impair
the performance of the device.
•
•
WaRnIng
WaRnIng
caUtIon
thE tUbE fIttIng may bE tIght -
do not
USE
oIlS oR gREaSE to lUbRIcatE tUbIng oR
oXygn oUtlEt poRt.
If tUbIng bEcomES dISconnEctEd dURIng op-
ERatIon, thE EclIpSE doES not alaRm. chEcK
floW at thE cannUla to vERIfy
oXygEn dElIvERy.
to EnSURE thERE IS adEQUatE floW to
dElIvER oXygEn, thE lEngth of thE tUbIng
mUSt not EXcEEd 50 fEEt (15.2m) WhEn US-
Ing contInUoUS floW modE, and mUSt not
EXcEEd 7 fEEt (2.1m) WhEn opERatIng In pUlSE
floW modE foR adEQUatE InSpIRatoRy Ef-
foRt dEtEctIon.