Vacuubrand CVC 3000 Instructions For Use Manual

Manual is about: Vacuum

Summary of CVC 3000

  • Page 1

    Vacuum controller cvc 3000 vacuum controller instructions for use original instructions n°: 999151_en_print technology for vacuum systems.

  • Page 2

    Imprint manufacturer: vacuubrand gmbh + co kg alfred‑zippe‑str. 4 97877 wertheim germany phone: head office +49 9342 808‑0 sales +49 9342 808‑5550 service +49 9342 808‑5660 fax: +49 9342 808‑5555 e‑mail: info@vacuubrand.Com web: www.Vacuubrand.Com 2 n°: 999151_en_cvc3000_print/20.07.2017 thank you f...

  • Page 3: 1   Introduction

    Content 3 n°: 999151_en_cvc3000_print/20.07.2017 1 introduction 7 1.1 user information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.2 about this document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2.1 display conventions . . . . . . . . . . . . . . . . . ...

  • Page 4: 6   Operation

    Content 4 n°: 999151_en_cvc3000_print/20.07.2017 5 operating and display elements 34 5.1 operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.1.1 selection knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.1.2 control panel . . . . . . . . . . . . . . ....

  • Page 5: 8   Resolving Problems

    Content 5 n°: 999151_en_cvc3000_print/20.07.2017 7.3 configuration menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 7.3.1 content selection . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 7.3.2 submenus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 7.4...

  • Page 6

    6 n°: 999151_en_cvc3000_print/20.07.2017 first steps (delivery status) first steps first steps on delivery status select language and units.

  • Page 7: 1   Introduction

    Introduction 7 n°: 999151_en_cvc3000_print/20.07.2017 1 introduction this manual is part of your product. It provides important instruc‑ tions for safe use of the product. Read this manual completely in order to understand proper use of your product. 1.1 user information safety read this manual thor...

  • Page 8: Contact

    Introduction 8 n°: 999151_en_cvc3000_print/20.07.2017 contact please ask for replacement in case of an incomplete manual or download instructions for use on our web page: www.Vacuubrand.Com contact us regarding any questions about this product, if you need further information, or to provide us with ...

  • Page 9: Notice

    Introduction 9 n°: 999151_en_cvc3000_print/20.07.2017 notice notice for a potentially harmful situation. Disregarding the notice could lead to material damage. Additional notes > information or specific use recommendation, which must be observed. > important information for the proper operation. > >...

  • Page 10: Additional Icons

    Introduction 10 n°: 999151_en_cvc3000_print/20.07.2017 additional icons positive example – do! Result – o. K. Negative example – do not! Refers to content of this manual. Refers to content of other documents. Press selection knob or key. Turn selection knob push and hold key push and turn selection ...

  • Page 11: 1.2.4  Abbreviations

    Introduction 11 n°: 999151_en_cvc3000_print/20.07.2017 1.2.4 abbreviations abs. Absolute atm standard atmospheric pressure cvc 3000 vakuum controller, controller d i (di) interior diameter dn nominal diameter ` atex equipment labeling ek emission condenser ex* outlet fkm fluorelastomer gr. Size hh:m...

  • Page 12: 2   Safety Instructions

    Safety instructions 12 n°: 999151_en_cvc3000_print/20.07.2017 2 safety instructions all safety instructions must be observed by all individuals working with the product described here. The safety instructions are valid for the complete life cycle of the product. 2.1 working conditions use the produc...

  • Page 13: Cvc 3000

    Safety instructions 13 n°: 999151_en_cvc3000_print/20.07.2017 controlling explosive atmoshpere, which does not compare to the atex approval of the cvc 3000  see rating plate. Unauthorized modifications an the use of accessories and spare parts that are not recommended by the manufacturer. Use in mi...

  • Page 14: 2.2  Target Groups

    Safety instructions 14 n°: 999151_en_cvc3000_print/20.07.2017 2.2 target groups ensure that the controller is only operated by authorized and skilled personnel. Users need to have the corresponding skills and qualifications for doing the job listed in the table user permissions . 2.2.1 user permissi...

  • Page 15: 2.3  Safety Precautions

    Safety instructions 15 n°: 999151_en_cvc3000_print/20.07.2017 2.3 safety precautions products of vacuubrand gmbh + co kg are subject to high quality tests with goals for safety and operation. Prior to delivery each product has been tested thoroughly. 2.3.1 protective clothing no special protective c...

  • Page 16: Notice

    Safety instructions 16 n°: 999151_en_cvc3000_print/20.07.2017 sources of error during connection notice measuring error due to an obstructed vacuum line. > > prevent overpressure > 1060 mbar (> 795 torr) in the piping system. Condensate can falsify the measurement. Position the vacuum hose in such a...

  • Page 17: Atex Marking

    Safety instructions 17 n°: 999151_en_cvc3000_print/20.07.2017 when using remote control please regard the following: > coordinate planned projects with colleagues sharing the equipment. > if necessary inform colleagues that you plan to use the remote control. > avoid different, parallel settings. In...

  • Page 18: 2.4  Safety and Service

    Safety instructions 18 n°: 999151_en_cvc3000_print/20.07.2017 2.4 safety and service safety regulations that apply to your work environment also apply to persons who perform service works, especially in the handling of hazardous materials. 2.4.1 meaning health and safety clearance products which are...

  • Page 19: Proofed and Confirmed.

    Safety instructions 19 n°: 999151_en_cvc3000_print/20.07.2017 5. Before returning the product, please send the signed health and safety clearance form to your local supplier or our service department. For all service works the safety clearance needs to be proofed and confirmed. > did your product ge...

  • Page 20: 3   Product Description

    Product description 20 n°: 999151_en_cvc3000_print/20.07.2017 3 product description goods arrival check the shipment for transport damage and completeness. > report any transit damage immediately to the supplier. Notice condensate can damage the controller. A large temperature difference between sto...

  • Page 21: Front Side

    Product description 21 n°: 999151_en_cvc3000_print/20.07.2017 front side 1 2 3 4 1 chemically resistant plastic housing 2 lc display 3 control panel and product name 4 selection knob rear side 9 5 7 10 6 11 8 11 5 ports for vacuu·bus ® components 6 mains connection ` power supply unit ` vario ® diap...

  • Page 22: 3.2  Functionality

    Product description 22 n°: 999151_en_cvc3000_print/20.07.2017 3.2 functionality the controller manages vacuum processes by controlling vacuum pumps, in‑line isolation‑ and/or air admittance valves. It controls process vacuum, cooling water and venting to demand. Valves and/or vacuum pumps are necess...

  • Page 23: 3.3  Operation Modes

    Product description 23 n°: 999151_en_cvc3000_print/20.07.2017 3.3 operation modes up to 5 different operation modes are selectable at the controller. Specific modifications can be realized by indivudual mode me‑ nus. Selectable operation modes pump down vac control program auto mode vacuulan for mor...

  • Page 24: Installation Conditions

    Installation and connection 24 n°: 999151_en_cvc3000_print/20.07.2017 4 installation and connection the controller is designed for installation directly at the workplace. > observe all specifications for installation, connection and operation according to technical data, see chapter 10.1.1 technical...

  • Page 25: 4.1.2  Built‑In Version

    Installation and connection 25 n°: 999151_en_cvc3000_print/20.07.2017 4.1.2 built‑in version the cvc 3000 can be used as built‑in version fixed with spring clips; e. G., as front controller of a vario ® pump, built into a cable duct cover or into the cut‑out of a switch cabinet. Cut‑out for switch c...

  • Page 26: 4.1.4  Examples of Use

    Installation and connection 26 n°: 999151_en_cvc3000_print/20.07.2017 example vario ® pump with cvc 3000 and rotary evaporator 4.1.4 examples of use 1 4 2 3 5 1 emission condenser with round flask 2 vario ® pump (pc 3001 vario ® pro) 3 controller cvc 3000, built‑in version 4 vacuum hose 5 rotary eva...

  • Page 27

    Installation and connection 27 n°: 999151_en_cvc3000_print/20.07.2017 2 4 7 6 3 5 1 1 wall power supply 2 controller cvc 3000, directly mounted 3 vacuu·bus ® cable 4 vacuum vessel, recipient 5 vacuum valve 6 vacuum hose 7 diaphragm pump, vacuum pump > > install the controller as close as possible to...

  • Page 28: 4.2  Connection

    Installation and connection 28 n°: 999151_en_cvc3000_print/20.07.2017 4.2 connection 4.2.1 electrical connection wall power supply kit* * short-circuit-proof multi-voltage power supply with integrated overload protection and changeable mains plugs. Prepare wall power supply plug 1. Take the wall pow...

  • Page 29: Connect to Mains

    Installation and connection 29 n°: 999151_en_cvc3000_print/20.07.2017 connect power supply to the controller > plug female connection of the power supply cable into mains connection of the controller. Consider new connection design: for easy connection, the cvc 3000 of the newest series have a guide...

  • Page 30: 4.2.2  Vacuum Connection

    Installation and connection 30 n°: 999151_en_cvc3000_print/20.07.2017 4.2.2 vacuum connection notice flexible vacuum hoses can contract because of evacuation. > > fix vacuum hose at the connections. > > fix connected components. > > measure and trim the vacuum hose to a length that cares for the max...

  • Page 31: Connection Examples

    Installation and connection 31 n°: 999151_en_cvc3000_print/20.07.2017 connection examples depending on design and installation the controller provides se‑ veral options for connection to the vacuum system. 3 1 3 2 flexible caoutchouc hose (1) or (2), directly plugged on the hose nozzle (3). 4 5 6 7 ...

  • Page 32: 4.2.3  Vacuu·bus

    Installation and connection 32 n°: 999151_en_cvc3000_print/20.07.2017 4.2.3 vacuu·bus ® vacuu·bus ® is a system for communication to peripheral ac‑ cessories which can be directly connected to the controller. The‑ se accessory components are self‑configuring by switching on the controller. All vacuu...

  • Page 33: Vacuu·bus

    Installation and connection 33 n°: 999151_en_cvc3000_print/20.07.2017 vacuu·bus ® components vacuum sensor vsk 3000 636657 vsp 3000 636163 vacuum gauge vacuu·view 683220 vacuu·view extended 683210 vacuum valve vv‑b 6 674290 (in-line valve) vv‑b 6c 674291 vv‑b 15c, kf 16 674210 vv‑b 15c, kf 25 674215...

  • Page 34: 5.1  Operating Elements

    Operating and display elements 34 n°: 999151_en_cvc3000_print/20.07.2017 5 operating and display elements 5.1 operating elements the operating elements are located on the controller front. Controller figure see chapter 3.1 vacuum controller cvc 3000 on page 20 5.1.1 selection knob the selection knob...

  • Page 35: 5.1.3  Key Combinations

    Operating and display elements 35 n°: 999151_en_cvc3000_print/20.07.2017 mode – select operation mode ` with stopped operation: mode menu for selecting the operation mode. Mode – change function/mode ` during running operation: to switch from pump down to vac control and further to auto mode . ` dur...

  • Page 36: Bar Graphic

    Operating and display elements 36 n°: 999151_en_cvc3000_print/20.07.2017 5.2 display and user interface after booting the pressure display appears, including bar graphic and preset operation mode. 5.2.1 pressure display 1 2 3 4 1 title bar (or status bar) operation mode ‑ mode ` pump down ` vac cont...

  • Page 37: 5.2.2  Display Icons

    Operating and display elements 37 n°: 999151_en_cvc3000_print/20.07.2017 5.2.2 display icons when vacuum control has started additional icons appear on the display. When starting operation icon meaning vacuum control is running (animation) process time; runtime vacuum control (hh:mm:ss) active compo...

  • Page 38: Additional Information

    Operating and display elements 38 n°: 999151_en_cvc3000_print/20.07.2017 status display while operation is running icon meaning pump down – continous pumping pump down: lower pressure limit reached vacuu·lan: pump down to set pressure vac control: for 2‑point control – pump down to set pressure vacu...

  • Page 39: Configuration/display

    Operating and display elements 39 n°: 999151_en_cvc3000_print/20.07.2017 error display active connection to rs232 port audio signal (beep) fault indication (warning symbol) icon meaning flashing: warning! When device connected to sub‑d (option) icon meaning controller in remote mode; controller only...

  • Page 40: Configuration

    Operating and display elements 40 n°: 999151_en_cvc3000_print/20.07.2017 5.2.4 menu display in general the controller includes several menus and submenus, e. G., configuration , function , display... . Submenu 1 3 2 1 titel bar ` menu name 2 menu interface ` selectable functions or submenus (on the ...

  • Page 41: 5.3  Handling Cvc 3000

    Operating and display elements 41 n°: 999151_en_cvc3000_print/20.07.2017 handling and operation 5.3 handling cvc 3000 the handling of the controller is menu‑driven. Menus are ac‑ cessed via push buttons on the control panel or via key combina‑ tions. Use the selection knob to select function or menu...

  • Page 42: Input (Data Entry)

    Operating and display elements 42 n°: 999151_en_cvc3000_print/20.07.2017 input (data entry) changeable values are positioned on the right side in the display. Text on the ride side accords to content selection like in a drop down list. Exception: menu program , in this menu data and value are editab...

  • Page 43: 6   Operation

    Operation 43 n°: 999151_en_cvc3000_print/20.07.2017 6 operation 6.1 switch‑on/‑off controller switch‑on cvc 3000 v2.Xx 5 initial screen: company logo and firmware version, for approximately 2 seconds. 5 pressure graphic is displayed. Switch‑off > press key on/off 5 controller switched off (display o...

  • Page 44

    Operation 44 n°: 999151_en_cvc3000_print/20.07.2017 6.2 select language and pressure unit 櫓匡 櫓匡 cvc 3000 v2.Xx select language and pressure unit 5 initial screen..

  • Page 45: Mode

    Operation 45 n°: 999151_en_cvc3000_print/20.07.2017 6.3 mode – operation mode the controller is supplied with several operating modes. Only when the controlling process is stopped it is possible to select an operation mode. 6.3.1 select operation mode 5 title bar shows the selected operation mode ( ...

  • Page 46: 6.3.2  Mode Menu

    Operation 46 n°: 999151_en_cvc3000_print/20.07.2017 6.3.2 mode menu mode menu display 1 3 2 1 title bar menu name 2 selectable operation modes pump down ` continous pump down or ` pump down with pressure and time presetting. ` vario ® : pump down with adjustable motor speed (pumping speed) and conti...

  • Page 47: 6.4  Start Controlling

    Operation 47 n°: 999151_en_cvc3000_print/20.07.2017 6.4 start controlling start vacuum controlling after selecting the required operation mode. The controller works in delivery status with the default set‑ tings of the factory setting. Start controlling 5 controller starts. 5 icons are displayed. 6....

  • Page 48: Venting

    Operation 48 n°: 999151_en_cvc3000_print/20.07.2017 venting the vent button is used to vent the system. A short click on this button will momentarily vent the system as the process contin‑ ues. Holding the vent key for longer than 2 seconds will cause the system to be vented to atmospheric pressure ...

  • Page 49: Mode

    Operation 49 n°: 999151_en_cvc3000_print/20.07.2017 6.5.2 change operation mode during running operation the operation mode can be switchted between pump down , vac control and auto mode 1 by pressing mode key. Switch mode during running operation 5 operation mode switched to vac control . 5 operati...

  • Page 50: Vac Control

    Operation 50 n°: 999151_en_cvc3000_print/20.07.2017 switch back from vac control to pump down . 45 ° 45 ° 5 title bar displays pump down . Typical applications pump down vac control : semiautomatic distillation. Recommended for applications for which the process vacuum is still to be determined. Fir...

  • Page 51: Graphic

    Operation 51 n°: 999151_en_cvc3000_print/20.07.2017 6.5.3 display graphic (curve) in addition to the bar graphic of the pressure display the display can be switched to a diagram named graphic which shows a pressure vs. Time curve. That graphic curve will only be displayed while operation is run‑ nin...

  • Page 52: Graphic Menu

    Operation 52 n°: 999151_en_cvc3000_print/20.07.2017 graphic menu 3 4 2 1 1 header ` active operation mode. ` vacuum set point; for vac control or auto mode . ` actual vacuum value (actual pressure). ` eelapsed process time. 2 axis – pressure ` unit according to pre‑settings (mbar, torr, hpa) 3 press...

  • Page 53: 6.6.1  Set Vacuum

    Operation 53 n°: 999151_en_cvc3000_print/20.07.2017 6.6 quick adaption during operation 6.6.1 set vacuum in mode vac control the set vacuum can be adapted directly du‑ ring running operation. Adapt set vacuum  fine tuning 1 detent = 1 pressure value (mbar, torr, hpa) 5 controller controls to new se...

  • Page 54: Pump Down

    Operation 54 n°: 999151_en_cvc3000_print/20.07.2017 6.6.2 motor speed (only vario ® ) in mode pump down the motor speed of a vario ® pump can be adjusted directly during running operation. Pump down – adjust motor speed 5 pump icon with percentage value. 5 vario ® pump runs with adjusted motor speed...

  • Page 55: 7.1  Operation Menus

    Advanced menus and operation 55 n°: 999151_en_cvc3000_print/20.07.2017 7 advanced menus and operation 7.1 operation menus a selected operation mode can be adapted and optimized for the process through the corresponding operation menu. The settings in an operation menu include mainly: motor speed, se...

  • Page 56: 7.1.1  Pump Down

    Advanced menus and operation 56 n°: 999151_en_cvc3000_print/20.07.2017 7.1.1 pump down continous pump down with pressure and time presettings. Menu – pump down parameter meaning speed (%) vario ® : speed settings for pump down. Adjustment range: 1–100; hi* minimum** (mbar, torr, hpa) vacuum set poin...

  • Page 57: 7.1.2  Vac Control

    Advanced menus and operation 57 n°: 999151_en_cvc3000_print/20.07.2017 7.1.2 vac control control to a set vacuum value. Menu – vac control parameter meaning set vacuum (mbar) setting for lower vacuum level for 2‑point control or precisely for vario ® pump. Adjustment range: turbo*; 1–1060 speed (%) ...

  • Page 58: Hysteresis Values Auto

    Advanced menus and operation 58 n°: 999151_en_cvc3000_print/20.07.2017 hysteresis values auto set vacuum (mbar) 5 10 50 80 100 200 500 700 900 1000 hysteresis (mbar) 2 2 5 8 9 17 40 55 71 78 7.1.3 auto mode (only valid for vario ® pump) automatic detection and tracking of boiling point, unaffected b...

  • Page 59: 7.1.4  Program

    Advanced menus and operation 59 n°: 999151_en_cvc3000_print/20.07.2017 application example – rotary evaporator use minimum to prevent rotary evaporator from receiving re‑eva‑ poration from the flask. Set minimum value to the vapor pressure of the solvent at ambient temperature. 7.1.4 program up to 1...

  • Page 60: 7.1.5  Vacuulan

    Advanced menus and operation 60 n°: 999151_en_cvc3000_print/20.07.2017 7.1.5 vacuulan vacuum control, optimized for vacuu·lan ® vacuum networks by vacuubrand. Menu vacuulan parameter meaning set vacuum (mbar) setting for lower vacuum level, which shall be reached easily when the vacuum network is no...

  • Page 61: 7.3  Configuration Menu

    Advanced menus and operation 61 n°: 999151_en_cvc3000_print/20.07.2017 7.3 configuration menu in menu configuration the controller parameters are set. This menu is also for adjusting the vacuum sensor and for loading defaults settings. Call up configuration menu 5 configuration menu displayed. After...

  • Page 62: 7.3.1  Content Selection

    Advanced menus and operation 62 n°: 999151_en_cvc3000_print/20.07.2017 7.3.1 content selection the following menu items of configuration can be selected, ac‑ tivated and used. Defaults selection meaning cancel leave menu item without default setting. Load load default settings. If defaults load is a...

  • Page 63: Adjustment

    Advanced menus and operation 63 n°: 999151_en_cvc3000_print/20.07.2017 adjustment selection meaning 1060–700 20–0 adjustment range of a vacuum sensor, internal or ex‑ ternal at atmospheric pressure (1060–700) or under vacuum (20– ~0). For further descriptions about sensor adjustment see chapter 9.2 ...

  • Page 64: Submenu – Sensors

    Advanced menus and operation 64 n°: 999151_en_cvc3000_print/20.07.2017 submenu – sensors in submenu sensors all connected sensors are listed. The inter‑ nal sensor is generally displayed as sensor . External sensors are listed with sensor type name and address. Display meaning inverse sensor = curre...

  • Page 65: Submenu – Rs‑232

    Advanced menus and operation 65 n°: 999151_en_cvc3000_print/20.07.2017 submenu – rs‑232 submenu rs-232 is applied for interface configuration, parame‑ ter adjustments and commands. See also online instructions for use . Parameter selection meaning baud 19200 default setting for transmission speed. T...

  • Page 66

    Advanced menus and operation 66 n°: 999151_en_cvc3000_print/20.07.2017 7.4 function menu (see online version of cvc manual) 7.5 program functions (see online version of cvc manual) detailed descriptions about function menu with address assign‑ ment or about differential pressure measurment see onlin...

  • Page 67: 8   Resolving Problems

    Resolving problems 67 n°: 999151_en_cvc3000_print/20.07.2017 8 resolving problems technical support > to identify errors and potential remedies, please refer to the troubleshooting table: fault – cause – remedy on page 69 in case you need additional assistance, please contact our service department....

  • Page 68

    Resolving problems 68 n°: 999151_en_cvc3000_print/20.07.2017 combinations of flashing display icons icon flash rate fault and meaning beep when sound on ` limit pressure reached 1x ` overpressure 1x ` process time elapsed 1x + ` venting valve 2x + ` in‑line suction valve 3x + ` coolant valve 4x ` ex...

  • Page 69: Fault

    Resolving problems 69 n°: 999151_en_cvc3000_print/20.07.2017 8.2 fault – cause – remedy fault ` possible cause 3 remedy personnel sensitive process not controllable ` motor speed too high. ` pumping speed too high. 3 reduce motor speed. User, specialist frequent error messa‑ ges of connected compo‑ ...

  • Page 70: Fault

    Resolving problems 70 n°: 999151_en_cvc3000_print/20.07.2017 fault ` possible cause 3 remedy personnel in‑line suction valve icon flashes ` in‑line suction valve removed. ` plug disconnected. ` in‑line suction valve defective. 3 check the connection. 3 check plug connection. 3 replace defective part...

  • Page 71: Fault

    Resolving problems 71 n°: 999151_en_cvc3000_print/20.07.2017 fault ` possible cause 3 remedy personnel no display ` controller switched off. ` power supply disconnec‑ ted. ` power supply not cor‑ rectly connected. ` mains voltage failure. ` controller defective ` cable break. 3 switch on the control...

  • Page 72: Fault

    Resolving problems 72 n°: 999151_en_cvc3000_print/20.07.2017 fault ` possible cause 3 remedy personnel sensors submenu is per‑ manently displayed ` submenu sensors... Does not automatically switch back to previous display. 3 select the required sensor by turning and pressing selection knob. User, sp...

  • Page 73: 8.3  Controller Reset

    Resolving problems 73 n°: 999151_en_cvc3000_print/20.07.2017 8.3 controller reset auto reset the following error indications will be reset automatically with remedy: overpressure process time elapsed limit pressure reached error air admittance valve error peltronic active reset several error indicat...

  • Page 74: Load Default Settings

    Resolving problems 74 n°: 999151_en_cvc3000_print/20.07.2017 > > note the settings of stored programs, before loading default . Load default settings important! Load factory settings.

  • Page 75

    Resolving problems 75 n°: 999151_en_cvc3000_print/20.07.2017 8.4 error of external components error messages for defective external components such as in‑ line suction valve, vacuum sensor, etc. Cannot be reset. > replace defective accessories or > send defective accessories for repair to your local...

  • Page 76: 9.1  Cleaning

    Cleaning and maintenance 76 n°: 999151_en_cvc3000_print/20.07.2017 9 cleaning and maintenance 9.1 cleaning this chapter does not contain descriptions for the deconta‑ mination of the controller. This chapter describes only simple cleaning and care measures. 9.1.1 controller clean surface > clean soi...

  • Page 77: 9.2  Sensor Readjustment

    Cleaning and maintenance 77 n°: 999151_en_cvc3000_print/20.07.2017 9.2 sensor readjustment notice for readjustment the reference pressures need to be known with certainty in the pressure range 20 – 700 mbar (15 – 525 torr) no adjustment is possible. > > check the accuracy of the pressure sensor in c...

  • Page 78: Adjustment Under Vacuum

    Cleaning and maintenance 78 n°: 999151_en_cvc3000_print/20.07.2017 adjustment under vacuum an adjustment under vacuum is only possible if the pressure is lower than 1. Evacuate the measurement connection of the controller or in case the connected external vacuum sensor vsk 3000 to a pressure adjustm...

  • Page 79: 2. Call Up Menu

    Cleaning and maintenance 79 n°: 999151_en_cvc3000_print/20.07.2017 adjustment at a reference pressure instead of adjustment under vacuum to a pressure ( within the range of 0 – 20 mbar (0 – 15 torr) is possible. 1. Evacuate the measurement connection of the controller or in case the connected extern...

  • Page 80: 10  Appendix

    Appendix 80 n°: 999151_en_cvc3000_print/20.07.2017 10 appendix 10.1 technical information product vacuum controller vacuum controller cvc 3000 internal vacuum sensor ceramic diaphragm (alumina), capacitive, gas independent, absolute pressure 10.1.1 technical data ambient conditions (us) working temp...

  • Page 81

    Appendix 81 n°: 999151_en_cvc3000_print/20.07.2017 vacuum data cvc 3000, internal vacuum sensor (us) measuring range, absolute 1080–0,1 mbar 810–0.1 torr max. Control range 1060–0,1 mbar 795–0.1 torr resolution 0,1 mbar 0.1 torr max. Admissible media temperature (gas): temporarily 80 °c 176°f contin...

  • Page 82

    Appendix 82 n°: 999151_en_cvc3000_print/20.07.2017 display type lc display (lcd) brightness control yes pressure display switchable: mbar, torr, hpa weight and dimensions* (us) weight (built‑in) 440 g 0.97 lb weight with foot (table top) 570 g 1.3 lb dimensions 123 mm x 124 mm x 83 mm 5 in. X 5 in. ...

  • Page 83: 10.1.3 Rating Plate

    Appendix 83 n°: 999151_en_cvc3000_print/20.07.2017 rating plate 10.1.3 rating plate > > in case of malfunction, please note type and serial number on the rating plate. > > when contacting our service department, name us product type and serial number. With this information we can offer selective sup...

  • Page 84: Sub‑D Panel Connector

    Appendix 84 n°: 999151_en_cvc3000_print/20.07.2017 10.2 interface commands (see online version of cvc manual) detailed descriptions about the complete interface commands see online instructions for use 999151 . 10.2.1 pin assignment (rs232) sub‑d panel connector sub‑d 9 poles (rear side of cvc 3000)...

  • Page 85: 10.3  Ordering Information

    Appendix 85 n°: 999151_en_cvc3000_print/20.07.2017 10.3 ordering information vacuum controller order‑n° cvc 3000 table top version 683160 cvc 3000 built‑in version 636595 cvc 3000 + vsp 3000 (package for fine vacuum control) 635983 spare parts order‑n° vacuum sensor (vacuum gauge head) vsk 3000, 108...

  • Page 86: Source of Supply

    Appendix 86 n°: 999151_en_cvc3000_print/20.07.2017 accessories order‑n° selection knob 612091 plug‑in rubber foot 638901 spring clip 636782 wall plug power supply 30w 24v; interchangeable mains plugs 612090 hose nozzle 636045 hose 636046 locking ring 10 mm for knurled nut m14 x 1 (637657) 637658 rou...

  • Page 87: 10.4  Service

    Appendix 87 n°: 999151_en_cvc3000_print/20.07.2017 10.4 service take advantage of the comprehensive service range of vacuubrand gmbh + co kg. Service in detail product guidance and practical solutions, fast delivery of spare parts and accessories, professional maintenance, immediate repairs processi...

  • Page 88: 10.5  Glossary

    Appendix 88 n°: 999151_en_cvc3000_print/20.07.2017 10.5 glossary hi mode scores maximum pumping speed and low ultimate vacuum with the pump and optimum speed for the respective pressure (auto‑ matic speed reduction at ultimate vacuum). Hysteresis regulates control performance of 2‑point control in m...

  • Page 89: 10.6  Index

    Appendix 89 n°: 999151_en_cvc3000_print/20.07.2017 10.6 index a action step . . . . . . . . . . . . . . . 10 active reset. . . . . . . . . . . . . . . 73 adapt set vacuum. . . . . . . . . . 53 additional notes . . . . . . . . . . . . 9 adjustment . . . . . . . . . . . . . . . 63 adjustment at atmosp...

  • Page 90

    Appendix 90 n°: 999151_en_cvc3000_print/20.07.2017 u user permissions . . . . . . . . . . 14 v vac control . . . . . . . . . . . . . . . 46 vacuu•bus. . . . . . . . . . . . . . 32 vacuu•bus accessories. . . . 33 vacuu•bus components . . . 33 vacuu•bus principle . . . . . . 32 vacuu•control . . . . ....

  • Page 91: 10.7  Certifications

    Appendix 91 n°: 999151_en_cvc3000_print/20.07.2017 10.7 certifications 10.7.1 ec declaration of conformity eu‑konformitätserklärung ec declaration of conformity déclaration ce de conformité hersteller / manufacturer / fabricant: vacuubrand gmbh + co kg · alfred‑zippe‑str. 4 · 97877 wertheim · german...

  • Page 92: 10.7.2 Us/can Certificate

    Appendix 92 n°: 999151_en_cvc3000_print/20.07.2017 10.7.2 us/can certificate.

  • Page 93

    Appendix 93 n°: 999151_en_cvc3000_print/20.07.2017 declaration of conformity – china rohs 2 version 01 of april 2017 / aw copyright 2017 declaration of conformity – china rohs 2 vacuubrand gmbh + co kg has made reasonable efforts to ensure that hazardous mate‑ rials and substances may not be used in...

  • Page 94

    Appendix 94 n°: 999151_en_cvc3000_print/20.07.2017 declaration of conformity – china rohs 2 version 01 of april 2017 / aw copyright 2017 注释: 此表格适用于所有产品。以上列出的元件或组件不一定都属于所附产品的组成。 note: table applies to all products. Some of the components or parts listed above may not be part of the enclosed product. ...

  • Page 95

    95 n°: 999151_en_cvc3000_print/20.07.2017.

  • Page 96

    Technology for vacuum systems manufacturer: vacuubrand gmbh + co kg alfred‑zippe‑str. 4 97877 wertheim germany phone: head office +49 9342 808‑0 sales +49 9342 808‑5550 service +49 9342 808‑5660 fax: +49 9342 808‑5555 e‑mail: info@vacuubrand.Com web: www.Vacuubrand.Com version : 999151_en_cvc3000_pr...