Vaisala HMP110 series User Manual - List Of Serial Commands
User's Guide _______________________________________________________________________
36 ___________________________________________________________________ M211060EN-G
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in
bold
type.
The notation
Enter
) key on
your computer keyboard. Press
Esc
to clear the command buffer before
starting to enter commands.
Table 10
List of Serial Commands (software version 2.0.7)
Command
Description
?
Output information about the device
AERR
Set analog output error level
AMODE
View or set the analog output mode
AOVER [ON/OFF]
Allow analog outputs to exceed their range
10%
ASEL
Set analog output parameters and scaling
CDATE
View or set the calibration date
CODE
View the order code of the probe
CRH
Calibrate and adjust RH measurement
CRHCLR
Clear adjustment of RH measurement
CT
Calibrate and adjust T measurement
CTCLR
Clear adjustment of T measurement
CTEXT
View or set the calibration information field
ERRS
List present probe errors
FILT [0.001 … 1]
Set the result filtering
FRESTORE
Restore factory settings
HELP
List available commands
INTV [0 ... 255 S/MIN/H]
Set the continuous output interval (for RUN
mode)
L
Displays user adjustment parameters
R
Start the continuous outputting
RESET
Reset the probe
RHLIMIT
Extend maximum RH reading
S
Stop the continuous outputting
SDELAY [0 … 255]
View or set serial line answer minimum
delay
SEND [0 ... 255]
Output the reading once
SNUM
View the serial number of the probe
UNIT
Select metric or non-metric output units
VERS
View software version of the probe
Summary of HMP110 series
Page 1
User's guide vaisala humidity and temperature probes hmp60 and hmp110 series m211060en-g.
Page 2
Published by vaisala oyj street address: vanha nurmijärventie 21, fi-01670 vantaa, finland mailing address: p.O. Box 26, fi-00421 helsinki, finland phone: +358 9 8949 1 fax: +358 9 8949 2227 visit our internet pages at www.Vaisala.Com © vaisala 2016 no part of this manual may be reproduced, publishe...
Page 3: Table Of Contents
_________________________________________________________________________________ vaisala _________________________________________________________________________ 1 table of contents chapter 1 general information ............................................................................ 5 about t...
Page 4
User's guide _______________________________________________________________________ 2 ____________________________________________________________________ m211060en-g operation ................................................................................................. 29 getting started ........
Page 5
_________________________________________________________________________________ vaisala _________________________________________________________________________ 3 adjustment procedure using serial line (hmp110, hmp110d and hmp113) ......................................................... 56 one-p...
Page 6: List Of Figures
User's guide _______________________________________________________________________ 4 ____________________________________________________________________ m211060en-g list of figures figure 1 hmp60 and hmp110 series probes ........................................ 11 figure 2 filters for hmp60, hmp1...
Page 7: General Information
Chapter 1 _________________________________________________________ general information vaisala _________________________________________________________________________ 5 chapter 1 general information this chapter provides general notes for the manual and the hmp60 and hmp110 series probes. About t...
Page 8: Version Information
User's guide _______________________________________________________________________ 6 ____________________________________________________________________ m211060en-g version information table 1 manual revisions manual code description m211060en-g january 2016. This manual. Modbus protocol added. U...
Page 9: Caution
Chapter 1 _________________________________________________________ general information vaisala _________________________________________________________________________ 7 caution caution warns you of a potential hazard. If you do not read and follow instructions carefully at this point, the product...
Page 10: Recycling
User's guide _______________________________________________________________________ 8 ____________________________________________________________________ m211060en-g recycling recycle all applicable material. Do not dispose of with regular household refuse. Regulatory compliances hmp60 and hmp110 ...
Page 11: Trademarks
Chapter 1 _________________________________________________________ general information vaisala _________________________________________________________________________ 9 trademarks vaisala intercap ® and vaisala humicap ® are registered trademarks of vaisala oyj. Windows® is a registered trademark...
Page 12: Product Overview
User's guide _______________________________________________________________________ 10 ___________________________________________________________________ m211060en-g chapter 2 product overview this chapter introduces the features and options of the hmp60 and hmp110 series probes. Introduction to h...
Page 13
Chapter 2 ___________________________________________________________ product overview vaisala ________________________________________________________________________ 11 1210-009 figure 1 hmp60 and hmp110 series probes no. Model description 1 hmp60 ip65 rated stainless steel probe with intercap® se...
Page 14
User's guide _______________________________________________________________________ 12 ___________________________________________________________________ m211060en-g basic features and options - analog and digital output options: - hmp60, hmp63, hmp110 and hmp113 analog output mode: two analog out...
Page 15: Filter Options
Chapter 2 ___________________________________________________________ product overview vaisala ________________________________________________________________________ 13 filter options for order codes, see section options and accessories on page 72. 1001-008 figure 2 filters for hmp60, hmp110, hmp1...
Page 16: Probe Mounting Clamp
User's guide _______________________________________________________________________ 14 ___________________________________________________________________ m211060en-g installation accessories (optional) for order codes, see section options and accessories on page 72. Probe mounting clamp the option...
Page 17
Chapter 2 ___________________________________________________________ product overview vaisala ________________________________________________________________________ 15 2. Slide the lower clamp part over to the bottom end of the upper part as shown in figure 6 below. 1209-002 figure 6 sliding the ...
Page 18: Probe Mounting Flange
User's guide _______________________________________________________________________ 16 ___________________________________________________________________ m211060en-g probe mounting flange the probe mounting flange is a silicone flange that can be used to hold the probe in a through-wall installati...
Page 19: Hmp110D And Hmp110T
Chapter 2 ___________________________________________________________ product overview vaisala ________________________________________________________________________ 17 duct installation kit for hmp60, hmp110, hmp110d and hmp110t the duct installation kit includes a plastic pipe with a flange (vai...
Page 20: Loop Power Converter
User's guide _______________________________________________________________________ 18 ___________________________________________________________________ m211060en-g loop power converter the loop power converter is an open frame module that converts one 0 ... 2.5 vdc voltage output to a 4 ... 20 m...
Page 21: Cables
Chapter 2 ___________________________________________________________ product overview vaisala ________________________________________________________________________ 19 cables connection cables have a straight, threaded female m8 connector on one end and open wires on the other end. Also other com...
Page 22: Installation
User's guide _______________________________________________________________________ 20 ___________________________________________________________________ m211060en-g chapter 3 installation this chapter provides you with information that is intended to help you install the hmp60 and hmp110 series p...
Page 23
Chapter 3 ________________________________________________________________ installation vaisala ________________________________________________________________________ 21 dimensions for hmp63 and hmp113 0912-103 figure 16 hmp63 and hmp113 dimensions 0912-102 figure 17 hmp63 and hmp113 with plastic ...
Page 24: Hmp110T Probes
User's guide _______________________________________________________________________ 22 ___________________________________________________________________ m211060en-g mounting the hmp60, hmp110, hmp110d and hmp110t probes hmp60, hmp110, hmp110d and hmp110t are designed to be mounted from the m12 th...
Page 25
Chapter 3 ________________________________________________________________ installation vaisala ________________________________________________________________________ 23 drilling instructions for duct installation kit 0505-178 figure 19 drilling instructions the following numbers refer to figure 1...
Page 26: Note
User's guide _______________________________________________________________________ 24 ___________________________________________________________________ m211060en-g mounting the hmp63 and hmp113 probes hmp63 and hmp113 probes do not have a thread on the probe body. For a convenient installation, ...
Page 27: Wiring
Chapter 3 ________________________________________________________________ installation vaisala ________________________________________________________________________ 25 wiring for a secure connection to the probe, connect to the 4-pin m8 connector using a threaded connector. The grounding method ...
Page 28
User's guide _______________________________________________________________________ 26 ___________________________________________________________________ m211060en-g table 5 pinout of the probe connector pin hmp60 / hmp63 / hmp110 / hmp113 / hmp110t hmp110d / hmp110ref wire color 1 5 ... 28 vdc (v...
Page 29
Chapter 3 ________________________________________________________________ installation vaisala ________________________________________________________________________ 27 wiring with the loop power converter to use the loop power converter module with a hmp60 or hmp110 series probe, make sure that:...
Page 30: Power Supply Requirements
User's guide _______________________________________________________________________ 28 ___________________________________________________________________ m211060en-g power supply requirements the operating voltage for the hmp60 and hmp110 series probes must be in the following range: table 6 opera...
Page 31: Operation
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 29 chapter 4 operation this chapter contains information that is needed to operate the hmp60 and hmp110 series probes. Getting starte...
Page 32: Note
User's guide _______________________________________________________________________ 30 ___________________________________________________________________ m211060en-g connecting to the serial interface the connection to the serial interface is via the 4-pin connector on the probe; see connector pin...
Page 33
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 31 for a description of the modes and the smode command that is used to change the mode, see section set serial interface mode on pag...
Page 34: Probes
User's guide _______________________________________________________________________ 32 ___________________________________________________________________ m211060en-g terminal application settings for digital probes the steps below describe how to connect to digital probes using the putty terminal ...
Page 35
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 33 0807-004 figure 23 putty terminal application.
Page 36: Modbus Mode
User's guide _______________________________________________________________________ 34 ___________________________________________________________________ m211060en-g accessing serial line command interface (rs-485 mode) from analog or modbus mode follow the steps below to connect to the serial lin...
Page 37: Modbus Communication
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 35 modbus communication the modbus variant used in hmp60 and hmp110 series probes is modbus rtu. For a list of the available modbus r...
Page 38: List Of Serial Commands
User's guide _______________________________________________________________________ 36 ___________________________________________________________________ m211060en-g list of serial commands all commands can be issued either in uppercase or lowercase. In the command examples, the keyboard input by ...
Page 39: View Device Information
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 37 table 11 additional commands for probes with rs-485 output command description ?? Output information about the device in poll mode...
Page 40: View Order Code
User's guide _______________________________________________________________________ 38 ___________________________________________________________________ m211060en-g view calibration information use the cdate command to view the calibration date and ctext to view the calibration info text. Date fo...
Page 41: View Serial Number
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 39 view serial number use the snum command to view the serial number of the probe. Snum example: snum serial number : h3640004 view s...
Page 42: Stop Measurement Output
User's guide _______________________________________________________________________ 40 ___________________________________________________________________ m211060en-g outputting the results continues in intervals issued with the command intv . You can stop the output by entering the s command. Stop...
Page 43: Set Serial Interface Mode
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 41 example (shows default settings): seri baud p d s : 19200 n 8 1 set serial interface mode use the smode command to set the operati...
Page 44: Note
User's guide _______________________________________________________________________ 42 ___________________________________________________________________ m211060en-g note in the run mode, the probe may send the measurement data message right as you are typing the s command to stop the sending. The...
Page 45: Set Probe Address
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 43 set probe address use the addr command to view or set the probe address. To operate in the poll mode, the probe must have an addre...
Page 46: Set Measurement Units
User's guide _______________________________________________________________________ 44 ___________________________________________________________________ m211060en-g set measurement units use the unit command to view or set the measurement units that are used in the serial line measurement message...
Page 47: Calibration Commands
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 45 calibration commands calibrate humidity measurement use the crh command to perform a one-point or two-point correction to the capa...
Page 48
User's guide _______________________________________________________________________ 46 ___________________________________________________________________ m211060en-g clear adjustment of rh measurement use the crhclr command to clear the adjustment of rh measurement that has been done using the crh...
Page 49
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 47 clear adjustment of t measurement use the ctclr command to clear the adjustment of temperature measurement that has been done usin...
Page 50: Other Commands
User's guide _______________________________________________________________________ 48 ___________________________________________________________________ m211060en-g other commands set analog output mode use the amode command to show or set the operation mode of the analog output. This command is ...
Page 51: Scaling
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 49 set analog output parameters and scaling use the asel command to show or set the output parameters and scaling of the analog outpu...
Page 52: Note
User's guide _______________________________________________________________________ 50 ___________________________________________________________________ m211060en-g set analog output error indication level if the device is malfunctioning, the analog output is set to a specified level. This overri...
Page 53: Extend Maximum Rh Reading
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 51 extend analog output range use the aover command to allow the analog output channels to exceed their specified range by 10%. The s...
Page 54: Display Command List
User's guide _______________________________________________________________________ 52 ___________________________________________________________________ m211060en-g example (on analog output, extend the maximum rh reading to 120% and scale the rh output on channel 2 from (0…1 v) 0...100% to (0…1 ...
Page 55: Reset The Probe
Chapter 4 _________________________________________________________________ operation vaisala ________________________________________________________________________ 53 close the connection in poll mode the close command closes the connection to the probe. Close example: close line closed reset the...
Page 56: Maintenance
User's guide _______________________________________________________________________ 54 ___________________________________________________________________ m211060en-g chapter 5 maintenance this chapter provides information that is needed in basic maintenance of the hmp60 and hmp110 series probes. P...
Page 57: Calibration Procedure
Chapter 5 _______________________________________________________________ maintenance vaisala ________________________________________________________________________ 55 new filters can be ordered from vaisala. For available filters, see section filter options on page 13. The order codes of the filt...
Page 58: Note
User's guide _______________________________________________________________________ 56 ___________________________________________________________________ m211060en-g adjustment procedure using serial line (hmp110, hmp110d and hmp113) note if you are adjusting a probe that is in the analog output m...
Page 59: Hmp113)
Chapter 5 _______________________________________________________________ maintenance vaisala ________________________________________________________________________ 57 - try the ?? Command in case the probe is in poll mode, open the line using open command if necessary. - check your serial line se...
Page 60
User's guide _______________________________________________________________________ 58 ___________________________________________________________________ m211060en-g 4. Wait for 20 – 40 minutes for the humidity to stabilize. 5. Start the terminal session and connect the usb cable to the probe. If ...
Page 61: Hmp110T)
Chapter 5 _______________________________________________________________ maintenance vaisala ________________________________________________________________________ 59 if you wish to remove the effects of rh calibration (returning the rh measurement of the probe to the factory calibrated state), s...
Page 62: Note
User's guide _______________________________________________________________________ 60 ___________________________________________________________________ m211060en-g 8. Use the errs command to see that no errors are active: errs 9. Give the ct command, with the temperature value of the reference a...
Page 63: Measurement
Chapter 5 _______________________________________________________________ maintenance vaisala ________________________________________________________________________ 61 - mi70 indicator. - connection cable for mi70 indicator (219980). - one humidity reference. One-point adjustment with a single off...
Page 64
User's guide _______________________________________________________________________ 62 ___________________________________________________________________ m211060en-g - one known and stable temperature reference. 1. Connect the probe to port i of the mi70 indicator. 2. Turn on the mi70 indicator. 3...
Page 65: Repair Maintenance
Chapter 5 _______________________________________________________________ maintenance vaisala ________________________________________________________________________ 63 repair maintenance changing the intercap® sensor (hmp60 and hmp63) this procedure restores the humidity measurement accuracy of th...
Page 66: Note
User's guide _______________________________________________________________________ 64 ___________________________________________________________________ m211060en-g changing the humicap® 180r sensor (hmp110, hmp110d and hmp113) note replacing the humidity sensor of the hmp110, hmp110d and hmp113 ...
Page 67: Troubleshooting
Chapter 6 ____________________________________________________________ troubleshooting vaisala ________________________________________________________________________ 65 chapter 6 troubleshooting this chapter describes common problems, their probable causes and remedies, and contact information for...
Page 68: Technical Support
User's guide _______________________________________________________________________ 66 ___________________________________________________________________ m211060en-g problem or message likely causes and solutions the probe is not responding to any serial commands. - disconnect the usb cable from t...
Page 69: Technical Data
Chapter 7 _____________________________________________________________ technical data vaisala ________________________________________________________________________ 67 chapter 7 technical data this chapter provides the technical data of the hmp60 and hmp110 series probes. Specifications performan...
Page 70: Relative Humidity
User's guide _______________________________________________________________________ 68 ___________________________________________________________________ m211060en-g performance (hmp110 and hmp110d) relative humidity measurement range 0 ... 100 %rh accuracy (incl. Non-linearity, hysteresis and rep...
Page 71: Performance (Hmp113)
Chapter 7 _____________________________________________________________ technical data vaisala ________________________________________________________________________ 69 performance (hmp113) relative humidity measurement range 0 ... 100 %rh accuracy (incl. Non-linearity, hysteresis and repeatabilit...
Page 72: Operating Environment
User's guide _______________________________________________________________________ 70 ___________________________________________________________________ m211060en-g operating environment operating temperature range hmp60 -40 ... +60 °c hmp63 -40 ... +60 °c hmp110 -40 ... +80 °c hmp113 -40 ... +60...
Page 73: And Hmp110T)
Chapter 7 _____________________________________________________________ technical data vaisala ________________________________________________________________________ 71 external loads 0 ... 1 r l min 10 k Ω 0 ... 2.5 / 0 ... 5 / 1 ... 5 v r l min 50 k Ω mechanics (hmp60, hmp110, hmp110d and hmp110...
Page 74: Options And Accessories
User's guide _______________________________________________________________________ 72 ___________________________________________________________________ m211060en-g options and accessories information on spare parts, accessories, and calibration products is available online at www.Vaisala.Com and...
Page 75: Modbus Reference
Appendix a _________________________________________________________ modbus reference vaisala ________________________________________________________________________ 73 appendix a modbus reference this appendix describes the modbus protocol implementation of the hmp60/hmp110 series probes. For info...
Page 76: Modbus Register Map
User's guide _______________________________________________________________________ 74 ___________________________________________________________________ m211060en-g modbus register map table 17 modbus register map (measurement values and settings) logical address (decimal) pdu address (hexadecima...
Page 77: Note
Appendix a _________________________________________________________ modbus reference vaisala ________________________________________________________________________ 75 pdu address actual address bytes used in a modbus protocol data unit. Lsw least significant word (bits 15 … 0). Msw most significa...
Page 78: Note
User's guide _______________________________________________________________________ 76 ___________________________________________________________________ m211060en-g device identification objects table 18 device identification objects object id (decimal) object id (hexadecimal) object name (as def...
Page 79
Www.Vaisala.Com *m211060en*.