Vaisala HMP110 series User Manual - General Information
Chapter 1 ________________________________________________________ General Information
VAISALA ________________________________________________________________________ 5
CHAPTER 1
GENERAL INFORMATION
This chapter provides general notes for the manual and the products.
About This Manual
This manual provides information for installing, operating, and
maintaining Vaisala Humidity and Temperature Probes HMP60,
HMP110 and HMP110T.
Contents of This Manual
This manual consists of the following chapters:
- Chapter
General Information, provides general notes for the manual
- Chapter 2, Product Overview, introduces the product features and
- Chapter 3, Installation, provides you with information that is intended
to help you install the product.
- Chapter 4, Operation, contains information that is needed to operate
- Chapter 5, Maintenance, provides information that is needed in basic
- Chapter 6, Troubleshooting, describes common problems, their
probable causes and remedies, and contact information for technical
- Chapter 7, Technical Data, provides the technical data of the product.
Summary of HMP110 series
Page 1
User's guide vaisala humidity and temperature probes hmp60, hmp110 and hmp110t m211060en-c.
Page 2
Published by vaisala oyj phone (int.): +358 9 8949 1 p.O. Box 26 fax: +358 9 8949 2227 fi-00421 helsinki finland visit our internet pages at www.Vaisala.Com © vaisala 2011 no part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may...
Page 3: Table of Contents
________________________________________________________________________________ vaisala ________________________________________________________________________ 1 table of contents chapter 1 general information ............................................................................ 5 about thi...
Page 4
User's guide ______________________________________________________________________ 2 ___________________________________________________________________ m211060en-c connecting to the serial interface .......................................25 installing the driver for the usb cable.....................
Page 5
________________________________________________________________________________ vaisala ________________________________________________________________________ 3 chapter 6 troubleshooting ................................................................................. 57 solving typical problems ...
Page 6: List of Figures
User's guide ______________________________________________________________________ 4 ___________________________________________________________________ m211060en-c list of figures figure 1 hmp60 (left) and hmp110 (right) ..........................................11 figure 2 filter options ...........
Page 7: General Information
Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 5 chapter 1 general information this chapter provides general notes for the manual and the products. About this manual this manual p...
Page 8: Version Information
User's guide ______________________________________________________________________ 6 ___________________________________________________________________ m211060en-c version information table 1 manual revisions manual code description m211060en-c june 2011. This manual. Applicable from software vers...
Page 9: Safety
Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 7 safety the product delivered to you has been tested for safety and approved as shipped from the factory. Note the following precau...
Page 10: Recycling
User's guide ______________________________________________________________________ 8 ___________________________________________________________________ m211060en-c recycling recycle all applicable material. Dispose of batteries and the unit according to statutory regulations. Do not dispose of wit...
Page 11: Trademarks
Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 9 trademarks vaisala intercap ® and vaisala humicap ® are registered trademarks of vaisala oyj. Windows® is a registered trademark o...
Page 12
User's guide ______________________________________________________________________ 10 __________________________________________________________________ m211060en-c this page intentionally left blank..
Page 13: Product Overview
Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 11 chapter 2 product overview this chapter introduces the product features and options. Introduction to hmp60, hmp110, and hmp110t vai...
Page 14: Basic Features and Options
User's guide ______________________________________________________________________ 12 __________________________________________________________________ m211060en-c table 3 quantities measured by the hmp60 and hmp110 quantity abbreviation metric unit non metric unit relative humidity rh %rh %rh dew...
Page 15: Filter Options
Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 13 filter options the available filters for hmp60, hmp110 and hmp110t are shown in figure 2 below. For order codes, see section option...
Page 16: Probe Mounting Clamp
User's guide ______________________________________________________________________ 14 __________________________________________________________________ m211060en-c installation accessories (optional) probe mounting clamp the optional mounting clamp makes it easy to install the probe on the wall of...
Page 17: Duct Installation Kit
Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 15 duct installation kit the duct installation kit includes a plastic pipe with a flange (vaisala order code: 215619). To install the ...
Page 18: Loop Power Converter
User's guide ______________________________________________________________________ 16 __________________________________________________________________ m211060en-c loop power converter the loop power converter is an open frame module that converts one 0 ... 2.5 vdc voltage output to a 4 ... 20 ma ...
Page 19: Cables
Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 17 cables the probe cables have a standard m8 connector. Also other compatible m8 series cables can be used. 0910-135 figure 7 cable w...
Page 20
User's guide ______________________________________________________________________ 18 __________________________________________________________________ m211060en-c this page intentionally left blank..
Page 21: Installation
Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 19 chapter 3 installation this chapter provides you with information that is intended to help you install the product. Dimensions 091...
Page 22: Mounting The Probe
User's guide ______________________________________________________________________ 20 __________________________________________________________________ m211060en-c mounting the probe hmp60, hmp110 and hmp110t are designed to be mounted from the m12 thread on the probe body or from the smooth part ...
Page 23
Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 21 drilling instructions for duct installation kit 42 1 2 3 Ø24 Ø3. 2 0505-178 figure 12 drilling instructions the following numbers ...
Page 24: Wiring
User's guide ______________________________________________________________________ 22 __________________________________________________________________ m211060en-c wiring for a secure connection to the probe, connect to the 4-pin m8 connector using a threaded connector. Connect the shield if using...
Page 25
Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 23 wiring with the loop power converter to use the loop power converter module with hmp60, hmp110, or hmp110t, make sure that: - the ...
Page 26: Power Supply Requirements
User's guide ______________________________________________________________________ 24 __________________________________________________________________ m211060en-c power supply requirements the operating voltage for the hmp60, hmp110 and hmp110t probes must be in the following range: - 5 ... 28 vd...
Page 27: Operation
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 25 chapter 4 operation this chapter contains information that is needed to operate the product. Getting started when the probe is conne...
Page 28: Note
User's guide ______________________________________________________________________ 26 __________________________________________________________________ m211060en-c note the vaisala usb cable is not designed for permanent installation. When using the usb cable, no separate power unit is needed. The...
Page 29
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 27 installing the driver for the usb cable before taking the usb cable into use, you must install the provided usb driver on your pc. W...
Page 30: In Rs-485 Mode
User's guide ______________________________________________________________________ 28 __________________________________________________________________ m211060en-c terminal application settings for probes in rs-485 mode the steps below describe how to connect to the hmp60, hmp110 and hmp110t using...
Page 31
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 29 0810-070 figure 16 putty terminal application.
Page 32: In Analog Mode
User's guide ______________________________________________________________________ 30 __________________________________________________________________ m211060en-c terminal application settings for probes in analog mode follow the steps below to connect to the serial line on probes that have been ...
Page 33: List of Serial Commands
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 31 list of serial commands all commands can be issued either in uppercase or lowercase. In the command examples, the keyboard input by ...
Page 34: View Device Information
User's guide ______________________________________________________________________ 32 __________________________________________________________________ m211060en-c table 9 additional commands for probes with rs-485 output command description ?? Output information about the device in poll mode addr...
Page 35: View Calibration Information
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 33 view calibration information use the cdate command to view the calibration date, ctext to view the calibration info text, and adjd c...
Page 36: View Software Version
User's guide ______________________________________________________________________ 34 __________________________________________________________________ m211060en-c view software version use the vers command to display the software version of the probe. Vers example: vers hmp60 / 1.00.0 serial line...
Page 37: Set Serial Line Settings
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 35 output the measurement message once use the send command to output the measurement values once. If the probe is in poll mode and the...
Page 38: Set Serial Interface Mode
User's guide ______________________________________________________________________ 36 __________________________________________________________________ m211060en-c set serial interface mode smode [ xxx ] where xxx = stop, run, or poll in stop mode: outputting only when command is issued, any comma...
Page 39: Set Output Interval
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 37 set output interval use the intv command to show or set the output interval of the serial line measurement messages (applies when r ...
Page 40: Set Probe Address
User's guide ______________________________________________________________________ 38 __________________________________________________________________ m211060en-c set probe address use the addr command to view or set the probe address. To operate in the poll mode, the probe must have an address. ...
Page 41: Set Measurement Units
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 39 set measurement units use the unit command to view or set the measurement units that are used in the serial line measurement message...
Page 42: Calibration Commands
User's guide ______________________________________________________________________ 40 __________________________________________________________________ m211060en-c calibration commands calibrate humidity measurement use the crh command to perform a one-point or two-point correction to the capacita...
Page 43
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 41 clear adjustment of rh measurement use the crhclr command to clear the adjustment of rh measurement that has been done using the crh...
Page 44
User's guide ______________________________________________________________________ 42 __________________________________________________________________ m211060en-c clear adjustment of t measurement use the ctclr command to clear the adjustment of temperature measurement that has been done using th...
Page 45: Other Commands
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 43 other commands set analog output mode use the amode command to show or set the operation mode of the analog output. Amode [ch1] [ch2...
Page 46: Extend Analog Output Range
User's guide ______________________________________________________________________ 44 __________________________________________________________________ m211060en-c extend analog output range use the aover command to allow the analog output channels to exceed their specified range by 10%. The scali...
Page 47: Reset The Probe
Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 45 connect to the probe in poll mode use the open command to connect to a probe that is in poll mode. Open [ aa ] where aa = address (0...
Page 48: Restore Factory Settings
User's guide ______________________________________________________________________ 46 __________________________________________________________________ m211060en-c restore factory settings use the frestore command to restore the factory settings to the probe. All user settings, including the user-...
Page 49: Maintenance
Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 47 chapter 5 maintenance this chapter provides information that is needed in basic maintenance of the product. Periodic maintenance the...
Page 50: Changing The Filter
User's guide ______________________________________________________________________ 48 __________________________________________________________________ m211060en-c changing the filter the filter on the probe should be replaced when it is damaged or dirty. 1. Turn the filter counter-clockwise to lo...
Page 51: Note
Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 49 adjustment procedure (hmp110) note if you are adjusting a hmp110 or hmp110t probe that is in the analog output mode, see section ter...
Page 52: Two-Point Adjustment of Rh
User's guide ______________________________________________________________________ 50 __________________________________________________________________ m211060en-c 7. When your serial connection is working, use the l command to see the current user adjustment parameters. L 8. Use the errs command ...
Page 53
Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 51 6. Verify that the connection works by giving the ? Command: ? If the probe does not respond with device information: - disconnect t...
Page 54: (Hmp110 and Hmp110T)
User's guide ______________________________________________________________________ 52 __________________________________________________________________ m211060en-c one-point adjustment of t measurement (hmp110 and hmp110t) to perform a one-point adjustment to the temperature measurement of the hmp...
Page 55
Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 53 10. Check with the l command that the user adjustment parameters have changed. If you wish to remove the effects of t calibration (r...
Page 56: Repair Maintenance
User's guide ______________________________________________________________________ 54 __________________________________________________________________ m211060en-c repair maintenance changing the intercap® sensor (hmp60) this procedure restores the humidity measurement accuracy of the hmp60. No ad...
Page 57: (Hmp110)
Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 55 changing the humicap® 180r sensor (hmp110) note replacing the humidity sensor of the hmp110 is not necessary in normal operation. If...
Page 58
User's guide ______________________________________________________________________ 56 __________________________________________________________________ m211060en-c this page intentionally left blank..
Page 59: Troubleshooting
Chapter 6 ___________________________________________________________ troubleshooting vaisala _______________________________________________________________________ 57 chapter 6 troubleshooting this chapter describes common problems, their probable causes and remedies, and contact information for t...
Page 60: Technical Support
User's guide ______________________________________________________________________ 58 __________________________________________________________________ m211060en-c problem or message likely causes and solutions any check sum error is active. For example: - program flash check sum error - parameter...
Page 61: Technical Data
Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 59 chapter 7 technical data this chapter provides the technical data of the product. Specifications performance (hmp60) relative humid...
Page 62: Performance (Hmp110)
User's guide ______________________________________________________________________ 60 __________________________________________________________________ m211060en-c performance (hmp110) relative humidity measurement range 0 ... 100 %rh accuracy (incl. Non-linearity, hysteresis and repeatability) te...
Page 63: Operating Environment
Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 61 operating environment operating temperature range hmp60 -40 ... +60 °c hmp110 -40 ... +80 °c hmp110t -40 ... +80 °c operating humid...
Page 64: Options and Accessories
User's guide ______________________________________________________________________ 62 __________________________________________________________________ m211060en-c options and accessories table 11 hmp60 options and accessories description item code vaisala intercap® sensor 15778hm vaisala intercap...
Page 65
Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 63 table 13 hmp110t options and accessories description item code probe mounting flange 226061 probe mounting clamps, 10 pcs 226067 se...
Page 66
Www.Vaisala.Com *m211060en*.