Vega VEGASON 61 Quick setup manual - 6 Supplement
14
6 Supplement
VEGASON 61 • 4 … 20 mA/HART - four-wire
51856-EN-160712
6 Supplement
6.1 Technical data
Note for approved instruments
Thetechnicaldataintherespectivesafetyinstructionsarevalidforapprovedinstruments(e.g.with
Exapproval).Thesedatacandifferfromthedatalistedherein,forexampleregardingtheprocess
conditions or the voltage supply.
Input variable
Measuredvariable
distance between lower edge of the transducer and
product surface
Measuringrange
Ʋ Liquids
upto5m(16.4ft)
Ʋ Bulk solids
upto2m(6.562ft)
Deadzone
0.25m(0.82ft)
Electromechanical data - version IP 66/IP 67
Cable entry
Ʋ Double chamber housing
– 1xcableglandM20x1.5(cable:ø5…9mm),1x
blindplugM20x1.5;plugM12x1forVEGADIS61
(optional)
or:
– 1xclosingcap½NPT,1xblindplug½NPT,plug
M12x1forVEGADIS61(optional)
or:
– 1xplug(dependingontheversion),1xblindplug
M20x1.5;plugM12x1forVEGADIS61(optional)
Spring-loaded terminals for wire cross-
section
<2.5mm²(AWG14)
Voltage supply
Operating voltage
Ʋ Non-ExandEx-dinstrument
20…72VDC,20…253VAC,50/60Hz
Power consumption
max.4VA;2.1W