Velleman AVM360 Operation Manual - page 8
AVM360 GB
7
4.8 Diode test
1. Set the range selector at selected range position :
x 1K for 0 - 150 µA test,
x 100 for 0 - 1.5 mA test,
x 10 for 0 - 15 mA test,
x 1 for 0 - 150 mA test.
2. Connect the diode to the tester.
For I
F
(forward current) test, connect the "N" terminal of the tester to the
positive polarity of the diode and the "P" terminal to the negative polarity of the
diode. For I
R
(reverse current) test, reverse the connection.
3. Read I
F
or I
R
one the LI scale provided.
4. Read the linear (forward) voltage of the diode on the LV scale while testing I
F
or I
R
.
5. Specifications
5.1 DC Voltage
Ranges : 0.1 - 0.5 - 2.5 - 10 - 50 - 250 - 1000 V
Accuracy at FSD : 3
Sensitivity : 20 k
Ω
/ V
Extension : 25 kV (with HV probe extra)
5.2 AC Voltage
Ranges : 10 - 50 - 250 - 1000 V
Accuracy at FSD : 4
Sensitivity : 9 k
Ω
/ V
Decibelmeter : -10 to +22 dB
0 dB = 1mW / 600
Ω
5.3 DC Current
Ranges : 50 µA (at 0.1 VDC position) - 2.5 mA - 25 mA - 0.25 A
Accuracy at FSD : 3
Volt Drop : 250 mV