CALIBRATION
IMPLEMENTS
F.
Implements
Several implements are available for use with the FDX - see Table 2. A hook and
flat implement are included with FDX. All implements are attached to the American
Standard 10-32 FDX load shaft thread. Metric thread adapters are available.
F.1
Attaching the FDX Hook and Flat Implement
Calibrate:
• Apply full-scale weight matching the display - in lbf.
• Push
SELECT with weight applied and movement stopped.
• FDX displays applied weight indicating successful calibration.
FDX will reject a calibration with weights that are higher or lower than FDX
full-scale capacity. It is possible for FDX to accept a calibration with weights
that are close to, but do not match the full scale capacity of FDX. This
erroneous calibration will give inaccurate readings.
Calibration Accepted
• FDX displays the actual applied weight indicating successful calibration. If
uuuuu (under) or nnnnn (over) is displayed the weight is not accepted, or other
problems exist - see Calibration Rejected (below).
Verificationisrecommendedbyapplyingweightstocheckaccuracyat
various points of FDX range.
• To exit calibration, push
SELECT when FDX full-scale capacity is displayed or at
anytime during or after the five point weight test.
• FDX flashes
donE and returns to normal operation.
• Auto-Calibration is complete and FDX is ready for use.
Calibration Rejected
• If the weight is not accepted, the display shows
uuuuu (under) or nnnnn (over).
Verify weight applied matches the FDX capacity.
• If the weight used does not match FDX capacity, repeat E.3.1 Auto-Calibration
with an accurate full scale weight.
• If
uuuuu or nnnnn is displayed again after repeating Auto-Calibration, the FDX
may be defective or damaged. Call Wagner Instruments for instructions.