WAGNER FORCE TEN Operation Manual - page 6
FORCE TEN
™
FDX FORCE GAGE
10
FORCE TEN
™
FDX FORCE GAGE
11
B.
Keypad
B.1 Keypad Description
FDX is operated with four button keypad that controls all functions.
ON/OFF
• Turns FDX on and off.
• Self-tests display if held down when turning gage on.
• At turn-on, display flashes Force Cell capacity and
software version.
If there is no display or if low battery is indicated, the battery may be low or not
securelyconnected.ConnectingtheACPowerAdapterwillconfirmthis.
ZERO
•
Returns display to zero in non-peak and peak mode.
• Returns display to zero with tare force or weight applied to the
load shaft.
PEAK
• Activates peak mode and recalls peak readings.
UNITS
• Selects units of measurement: lbf, kgf, N, or ozf.
Eachofthebuttons,exceptON/OFF,alsohasset-upfunctions.
B.2 Keypad Control
Three black keypad buttons have the following additional functions:
SCROLL
• Review menu options.
• Review selections within each option.
SELECT
• Select displayed item.
• Save selection.
ESCAPE
• Cancel and return to previous menu option.
• Exit set-up menu.
Prior to actual test use, gain keypad and menu familiarity with practice.
KEYPAD
MENU OPERATIONS
C.
Menu Set-Up
C.1 Menu Description
FDX features options for selection in the set-up menu. At first level are options; at
second level are selections within the options:
TABLE 4 MENU OPTIONS AND SELECTIONS
Display
Option - Level One
Selections - Level Two
AoFF
Automatic Off
No* or 30 minutes
PSS
Peak Sampling Rate
• Sampling rate:
100 samples per second.*
Input bandwidth: 9Hz in non-peak
33Hz in peak.
• Sampling rate:
1000 samples per second.
Input bandwidth: 90Hz in non-peak
330Hz in peak.
L CAL
Auto - Calibration
See E.3 Auto-Calibration
* Factory default settings.
Inputbandwidth:Thebandoffrequenciesthatfilteringpassesthrufromtheloadcell
to the display. All other background noise or interference frequencies are excluded.