WAGNER FORCE TEN Operation Manual - page 7
FORCE TEN
™
FDX FORCE GAGE
12
FORCE TEN
™
FDX FORCE GAGE
13
C.2 Menu Operations
C.2.1 Menu Access
• Turn FDX off.
• Push / Hold
ZERO, push On/OFF.
• Release
On/OFF, then ZERO, AoFF is displayed.
C.2.2 Menu Instructions
• After entry into set-up menu, first option displayed is
AoFF.
•
SCROLL to review options: AoFF, PSS and LCAL.
•
SELECT to choose an option and move thru selections.
Currentsettingisdisplayed.PushSELECTtoretainthatsetting.
•
SCROLL to review selections - see Table 4.
•
SELECT to choose the selection: donE will appear on the display, previous
menu is displayed.
• Push
ESCAPE to exit after all selections are made or at anytime to exit menu.
Continuing to press SCROLL without pressing SELECT will move the display
thru the available options with no changes made to settings.
MENU OPERATIONS
MENU OPERATIONS
C.2.3 Menu Options
1)
AoFF Automatic-Off - Useful for battery power conservation.
FDX automatically shuts off after a period if no buttons are pressed:
No Operates continuously, no automatic-off, consumes more
battery power.
30 Shuts off after 30 minutes of inactivity, conserves battery power.
2)
PSS
Peak Sampling Rate and Filter.
FDX provides selectable peak sampling rates and corresponding filtering.
►
Routine Testing
Peak sampling rate of 100/sec is preferable for slowly occurring
events and significantly reduces battery power consumption.
►
Rapid Event Testing
Peak sampling rate of 1000/sec is required to capture peaks of
rapidly occurring events and break tests. The FDX takes 1000
“looks” per second to accurately capture a break point and
display it.
►
Digital Filtering
Filtering of displayed data provides accurate force readings by
eliminating vibrations and electromagnetic interference. Filtering is
simultaneously set with peak sampling rate as follows:
100 Samples/Second
• Routine Testing - Peak Off:
9 Hz Input Band Width
• Rapid Event Testing - Peak On:
33 Hz Input Band Width
1000 Samples Per Second
• Routine Testing - Peak Off:
90 Hz Input Band Width
• Rapid Event Testing - Peak On:
330 Hz Input Band Width
Inputbandwidth:Thebandoffrequenciesthatfilteringpassesthrufrom
the load cell to the display. All other background noise or interference
frequencies are excluded from the data display.
3)
L CAL
Auto - Calibration
FDX features “Smart Load Cell” calibration for verification and
adjustment without returning it to the factory. See E.