3Disc FireCR+ Service Manual - page 18
Manual is about: Computed Radiography Scanner
of
40
Service Manual
Fire
CR
+
Computed Radiography Scanner
18
Parts Removal/Replacement |
Fire
CR
+
4. Unscrew the 6 screws that hold down the board and replace (2.5mm hex).
5.
Make sure to connect the cables correctly when reassembling. The cables all have tags
corresponding to descriptions on the connectors.
6. Once you've finished the replacement of the main board, replace the internal and Top Cover.
Core Board
1. Remove Top Cover. (Pg. 14)
2. Remove Internal Cover. (Pg. 16)
3. Remove the Loctite on the nuts.
4. Unscrew the 4 nuts on the core board.