A&D UA-767PC Instruction Manual

Other manuals for UA-767PC: Instruction Manual, Instruction Manual
Manual is about: Digital blood pressure monitor

Summary of UA-767PC

  • Page 1

    Digital blood pressure monitor ua-767pc instruction manual 1wmpd4000034e.

  • Page 2: Contents

    2 contents dear customers .....................................................................................................3 preliminary remarks ..............................................................................................3 precautions...............................................

  • Page 3: Dear Customers

    3 dear customers congratulations on purchasing a state-of-the-art a&d blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manual caref...

  • Page 4: Parts Identification

    4 parts identification systolic display pressure switch diastolic display start button pulse display clock display data port air socket measurement in progress clock button memory access indicator low battery display 1.5v batteries (r6p, lr6 or aa) air connector plug arm cuff air hose.

  • Page 5: Symbols

    5 symbols symbols function / meaning treatment turns on or off the device sets the clock / does not store the current data direction guide to install battery direct current sn serial number date of manufacture type bf: device, cuff and tubing are designed to provide special protection against electr...

  • Page 6: Setting The Clock

    6 setting the clock the ua-767pc blood pressure monitor is equipped with a date and time function. To set the clock on the monitor, follow these steps: press the clock button on the front of the monitor. The blinking digits reflect the year. If they are not currently set on the correct year, press t...

  • Page 7: Preparation For Measurements

    7 left arm hose 2-3 cm (1 inch) excessive constriction cuff preparation for measurements 1. Installing / changing the batteries remove the battery cover and insert new batteries into the battery compartment as shown, taking care to observe the proper direction for each battery. Use r6p, lr6 or simil...

  • Page 8: Measurements

    8 measurements 1. Place the cuff on the arm (preferably the left arm). 2. Set the pressure switch to a value that will be 30 to 40 mmhg above the expected systolic value and press the start button. When the start button is pressed, all of the display symbols will appear for about one second. When th...

  • Page 9: Measurements

    9 measurements remain still while the device is in operation. When a measurement is made with insufficient pressure, the device will pump again to a higher pressure. 4. When measurement has been completed, the buzzer sounds and the air is automatically released from the cuff. The systolic pressure, ...

  • Page 10: Measurements

    10 measurements accurate measurement cannot be made if the exhaust velocity is outside the range of 2 to 5 mmhg. This can be determined by watching the display each time the "measurement in progress" symbol blinks. Note the reading as the "measurement in progress" symbol blinks. Notes for proper mea...

  • Page 11: Data In Memory

    11 data in memory storing data in memory this device can automatically store up to 280 sets of measurement data in its memory. If an error message appears, the measurement data will not be stored. Whether the data will be stored or not after the measurement can be selected as follows: when the start...

  • Page 12

    12 transmitting readings to a computer the connected computer is not permitted to be in the patient area. To transmit the readings to a computer, connect the attached communication cable to the data port on the front of the monitor. When the monitor receives the open command, it will be in the ready...

  • Page 13: About Blood Pressure

    13 about blood pressure what is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmhg). One's natu...

  • Page 14: Troubleshooting

    14 when recording your blood pressure. Consult your doctor to interpret your blood pressure data. Troubleshooting problem possible reason recommended action batteries are drained. Replace all batteries with new ones. Nothing appears in the display, even when the power is turned on. Battery terminals...

  • Page 15: Maintenance

    15 maintenance do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the a&d service group. The a&d service group will provide tech...

  • Page 16: A&d Co.,ltd.

    A&d co.,ltd. 1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d engineering, inc. 1756 automation parkway, san jose, ca. 95131 u.S.A. Telephone: [1] (408) 263-5333 fax: [1] (408)263-0119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon...