A&D UA-767PC Instruction Manual

Other manuals for UA-767PC: Instruction Manual, Instruction Manual
Manual is about: Digital blood pressure monitor

Summary of UA-767PC

  • Page 1

    Digital blood pressure monitor model ua-767pc instruction manual original manuel d ’ instructions traduction manual de instrucciones traducción manuale di istruzioni traduzione bedienungsanleitung Übersetzung gebruiksaanwijzing vertaling 使用手冊 1wmpd4002088b 翻譯.

  • Page 2: Contents

    English 1 contents dear customers .....................................................................................................2 preliminary remarks ..............................................................................................2 precautions.......................................

  • Page 3: Dear Customers

    English 2 dear customers congratulations on purchasing a state-of-the-art a&d blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this device will facilitate your daily blood pressure regimen. We recommend that you read through this manua...

  • Page 4: Contraindications

    English 3 contraindications the following are precautions for proper use of the device. Do not apply the cuff to an arm with another medical electrical equipment attached. The equipment may not function properly. People who have a severe circulatory deficit in the arm must consult a doctor before us...

  • Page 5: Parts Identification

    English 4 parts identification m systolic display pressure switch diastolic display pulse display clock display data port air socket measurement in progress clock button memory access indicator 1.5v batteries (r6p, lr6 or aa) low battery display dc jack start button proper fit range index mark arter...

  • Page 6: Symbols

    English 5 symbols symbols function / meaning treatment turns on or off the device sets the clock / does not store the current data direction guide to install battery direct current sn serial number date of manufacture type bf: device, cuff and tubing are designed to provide special protection agains...

  • Page 7: Setting The Clock

    English 6 setting the clock the ua-767pc blood pressure monitor is equipped with a date and time function. To set the clock on the device, follow these steps: press the clock button on the front of the device. The blinking digits reflect the year. If they are not currently set on the correct year, p...

  • Page 8: Preparation For Measurements

    English 7 preparation for measurements 1. Installing / changing the batteries remove the battery cover and insert new batteries into the battery compartment as shown, taking care to observe the proper direction for each battery. Use r6p, lr6, aa or similar type batteries. Caution insert the batterie...

  • Page 9: Preparation For Measurements

    English 8 preparation for measurements 4. Selecting the correct cuff size using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit rang...

  • Page 10: Preparation For Measurements

    English 9 preparation for measurements 5. Applying the arm cuff 1. Wrap the cuff around the upper arm, about 1-2 cm above the inside of the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse, and result in a measurement error. 2. Constriction of the upper ...

  • Page 11: Measurements

    English 10 measurements 1. Place the cuff on the arm (preferably the left arm). 2. Set the pressure switch to a value that will be 30 to 40 mmhg above the expected systolic value and press the start button. When the start button is pressed, all of the display symbols will appear for about one second...

  • Page 12: Measurements

    English 11 measurements remain still while the device is in operation. When a measurement is made with insufficient pressure, the device will pump again to a higher pressure. 4. When measurement has been completed, the buzzer sounds and the air is automatically released from the cuff. The systolic p...

  • Page 13: Measurements

    English 12 measurements accurate measurement cannot be made if the exhaust velocity is outside the range of 2 to 5 mmhg. This can be determined by watching the display each time the "measurement in progress" symbol blinks. Note the reading as the "measurement in progress" symbol blinks. Notes for pr...

  • Page 14: Data In Memory

    English 13 data in memory storing the data in memory this device can automatically store up to 280 sets of measurement data in its memory. If an error message appears, the measurement data will not be stored. Whether the data will be stored or not after the measurement can be selected as follows: wh...

  • Page 15

    English 14 transmitting readings to a computer the connected computer is not permitted to be in the patient area. To transmit the readings to a computer, connect the attached communication cable to the data port on the front of the device. When the device receives the open command, it will be in the...

  • Page 16: About Blood Pressure

    English 15 about blood pressure what is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmhg). On...

  • Page 17: Troubleshooting

    English 16 regular readings give a more comprehensive blood pressure history. Be sure to note date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data. Troubleshooting problem possible reason recommended action batteries are drained. Replace all bat...

  • Page 18: Maintenance

    English 17 maintenance do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the a&d service group. The a&d service group will prov...

  • Page 19

    English 18 applied part cuff type bf useful life device: 5 years (when used six times a day) cuff: 2 years (when used six times a day) accessory ac adapter the adapter is to connect the blood pressure monitor to a power source at home. Tb-233 please contact your local a&d dealer for purchasing. The ...

  • Page 20

    1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments limited unit 24/26 blacklands way, abingdon business park, abingdon, oxfordshire ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d engineering, inc....