A&D UA-767PC Instruction Manual - Troubleshooting
English 16
Regular readings give a more
comprehensive blood pressure history. Be
sure to note date and time when recording
your blood pressure. Consult your doctor to
interpret your blood pressure data.
Troubleshooting
Problem
Possible Reason
Recommended Action
Batteries are drained.
Replace all batteries with new ones.
Nothing appears
in the display,
even when the
power is turned
on.
Battery terminals are not
in the correct position.
Reinstall the batteries with negative
and positive terminals matching
those indicated on the battery
compartment.
The cuff does not
inflate.
Battery voltage is too low.
(LOW BATTERY
mark) blinks. If the
batteries are drained
completely, the mark
does not appear.
Replace all batteries with new ones.
The cuff is not applied
properly.
Apply the cuff correctly.
You moved your arm or
body during the
measurement.
Make sure you remain very still and
quiet during the measurement.
The cuff position is not
correct.
Sit comfortably and still.
Raise your hand so that the cuff is at
the same level as your heart.
The unit is not
operating
properly.
Readings are too
high or too low.
If you have a very weak or irregular
heat beat, the device may have
difficulty in determining your blood
pressure.
The value is different
from that measured at a
clinic or doctor’s office.
See “Why measure blood pressure
at home”.
Other
Remove the batteries. Replace
them properly and try the
measurement again.
Note: If the actions described above do not solve the problem, contact the
dealer. Do not attempt to open or repair this product, as any attempt to
do so will make your warranty invalid.