A&D UA-767PC Instruction Manual - Maintenance
Maintenance
Do not open the case of the device because it uses delicate electrical
components and an intricate air unit that could be damaged. If you can not locate
and fix the problem, request service from your supplier, or from the A&D service
group. The A&D service group will support authorized suppliers about technical
information, spare parts and units.
The technical testing procedures – to be carried out at least every 3 years – can
be performed either by the manufacturer or by an authorized repair service in
accordance with the regulations governing manufactures of medical products.
Specifications
Performance Specifications
Measurement method Oscillometric measurement
Measurement range Pressure : 20 - 280 mmHg
Pulse :
40 - 200 beats / minute
Accuracy
Pressure :±3 mmHg or 2%, whichever is greater
Pulse
:
±5%
Battery
4 x 1.5V (R6P, LR6, or AA)
Cuff Arm
Circumference
Classification
Type BF
0366
Device, cuff and tubing are designed to provide special
protection against electric shocks.
Clinical test
According to ANSI / AAMI SP-10 1987
Standard deviation
Mean error
Systolic blood pressure
5.43 mmHg
2.24 mmHg
Diastolic blood pressure
5.18 mmHg
1.54 mmHg
Environment Specifications
Operating environment
+10 °C - +40 °C, less than 85 %RH
Storage environment
−10 °C - +60 °C, less than 85 %RH
Physical Specifications
Dimensions
163.7 [W] x 111 [D] x 66.7 [H] mm
Weight
Approx. 320g without batteries
15