Abcon ProShip Instruction Manual - Calibration
10 PROSHIP
®
CALIBRATION
3. Press
[ZERO], the display will now say “CAL”andthena5digitnumberwhichisa
special internal reference number. The last number(s) on the right will change up or
downslowly.Whenthesenumbersarehardlychangingand/orthestableindicatoris
showing press
[UNIT].Thedisplaywillnowflash “- - - - -” forasecondandthenflash
with the weight you entered.
• Place the item(s) to be used for calibration on the platform, wait for a moment and
then press
[UNIT].Another5digitnumberwillappearonthedisplaywhichyouDO
NOTneedtoworryabout.
• Press
[ON/OFF] toturnscaleofftocompleteandsavethecalibration.
Ifamistakehasbeenmade,pressing
[ON/OFF] before the end of step 3, will turn the
scaleoffandcalibrationwillnothavetakenplace.
To test the calibration was successful, turn on the scale by pressing [
ON/OFF], once the
scale has finished self testing and the display shows
0, put item(s) used for calibration on
theplatformandcheckthescaleisweighingcorrectly.Ifnot,repeatsteps2&3again.