Animas Vibe Quick Reference Manual - Ezprime Screen:
ezPrime Screen:
Lets you properly align the cartridge and piston rod.
Also lets you prime your infusion set tubing and fill
your cannula or needle, before connecting to your
pump.
Prime/Rewind
Tips:
• A full Prime/Rewind sequence is required each time you change the battery.
• Your pump must be primed in order to access the BOlUS MEnU.
• The Fill Cannula step is not required for your pump to operate. For example,
when you prime your pump after a battery change and you are not inserting a new
infusion set, this step is not necessary.
ezPrime
Main Menu
Rewind
Load Cart
Prime
Fill Cannula
•
Rewind - Retract the piston to allow a new cartridge to be installed.
•
load Cart - Prepare the cartridge and piston to deliver insulin.
•
Prime - Remove air from the infusion set by filling it with insulin.
•
Fill Cannula - Remove air from the cannula by filling it with insulin.
m WARNING: Never prime the tubing or tighten
the cartridge cap while infusion set is connected
to your body. Failure to disconnect your infusion
set from your body before performing this step
can result in over delivery of insulin, and possible
injury or death. If your pump sustains internal
damage, the amount of unintended insulin delivery
could be significant. This could result in serious
injury or death from hypoglycemia.
9.
CAUTION - Investigational device. Limited by Federal law to investigational use.