Animas Vibe Quick Reference Manual - Bg Calibration
BG Calibration
BG Calibration Screen:
Lets you enter
fingerstick BG values for CGM
calibration.
When to calibrate
•
When the 2-hour CGM startup period ends.
– The pump will prompt you to enter BG values
from two separate fingersticks.
•
At least once every 12 hours.
– You may enter a fingerstick BG value at any time,
or the pump will prompt you if you forget.
•
If the fingerstick BG value you enter for startup calibration or calibration
update has not been accepted.
– The pump will prompt you to enter a fingerstick BG value, and will repeat the
prompt every 15 minutes until a value is accepted.
BG Calibrate
CALIBRATE
CANCEL
6.7
mmol/L
Tips:
• You must take a fingerstick test with your BG meter and use that BG value to make
any insulin or treatment decisions. Insulin dosing decisions should not be based
solely on results from the Dexcom G4 CGM Sensor and Transmitter.
• A BG value must meet the following criteria to be used for calibration:
– The BG test must have been taken within the last 5 minutes.
– The value must be within 2.2–22.2 mmol/l.
– The value must be from a fingerstick. Do Not use a BG test result from an
alternative sampling site (e.g., palm or forearm) for CGM calibration.
– A BG value entered when using the ezCarb and/or ezBG feature on the pump can
also be used for a CGM calibration.
15.
CAUTION - Investigational device. Limited by Federal law to investigational use.
41028551 RevA_QRG_Unity_CA_EN_mmolL_r1.indd 15