Bacsa ACS-Z User Manual - page 6
VERSIÓN 12.11.06
3.4. Calibration
Keep pressing【 Cali.】 for about 4 seconds. Calibrate after the weight window
displays “CAL.”. There are two ways to calibrate:
1. Full range load: Put enough weight on the platform according to the full scale.
Press【E nter】 to make sure. Calibration finished.
2. Random load:
2.1. Press【Set】 , the weight window displays the weight calibrated last time.
2.2. Press【 ↑
】 and【Enter】 to input the weight you want to load. Then load the
corresponding weight. Press【E nter】 to make sure. Calibration finished.
Press【Model】 to exit from the calibration state in the process of calibrating.
Note:The high limit and low limit set in the weighing model are saved in the scale.
Different unit corresponding to different alarm range and method. The value
inputted by pressing
【 ↑】
cannot be more than the full scale.
4. ALARM INDICATION:
1. The weight window displays “--OF--” with continuously alarm sound when the
weight over 100%FS+9d. If the ADC overflows, the weight window displays
“—Adc—”with continuously alarm sound. For these instance, the weight
should be taken away.
2. If the voltage of the accumulator is low, the weight window displays “—Lo—”
while the weight is zero (The weight display will recover to normal when it is
loaded ). Under this condition, you can use the scale in a short time, but you
should plug in the AC plug as soon as possible to recharge the accumulator.
3. The weight window displays “HHHH” or “LLLL” when the zero weight is higher
or lower than the permitted range.
4. When you switch on the scale. The weight window will display “UNSTA” if the
scale is not stable. It maybe because of the platform’s intensity is too low or
the platform is shaken strongly.
You may strengthen the platform or avoid the shake.
Balanza ASC