5. If it display “-SYS-” when you switch on the scale please re-calibrate or send it
to repair.
If it display –Set--”, it means the alarm setting is wrong, please reset the alarm
value.
5. RS232 TRANSMISSION METHOD:
5.1. RS232 setting
Press 【Mode l】 and 【 Tare】 key when self-testing, loosen two key when hear the
buzzer sound and enter the RS232 transmission setting.
5.1.1. Set transmission method
Press the 【 set】 key to choose the method of:
“Stb” = output when stable,
“Etb“ = output once when【 Enter】 key is pressed,
“SEr” = series (continuous output), or
“CLOSE” = output disabled.
Then press【 ↑
】 key to select method or press 【 Enter】 to confirm and set the next
parameter. (The default mode is“close”.)
5.1.2. Baud rate Setting
Press 【 ↑
】 key to select the desired baud rate among “1200”“2400”“4800”“9600”,
then press 【 Enter】 to confirm the selection and exit setting.
( The default baud ratio is 2400.)
Stb
Etb
SEr
CLOSE
↑
Transmit
continuously
Disable the
transmit
↑
↑
↑
print once
when stable
Press 【 Enter】
to send
9600
↑
4800
↑
2400
↑
1200
↑