Badger Meter 340BN Installation & Operation Manual - page 7
7
9.
Program the Model 340BN transmitter using the diagram
below as a reference.
Note #1:
Badger Meter sensors are pulse type sensors. The K and Offset
information is printed in the owners manual shipped with the prod-
uct. This information is also available on our website. Calibration
constants for other sensors must be supplied by the manufac-
turer.
Note #2
Typically the temperature measured by T1 will be greater than T2
in a heating application and less than T2 in a cooling application.
The selection of one of these choices will determine if energy
calculations are made for heating only (T1>T2), cooling only
(T1 Note #3 The filter coefficient screen allows adjustment of the flow and energy filters. Filter settings determine how the Series 340BN responds to actual changes in flow and energy. A scale of 0 -10 is used with 10 providing the greatest degree of smoothing. Step 1 Select the flow sensor type and enter the K and Offset values. See Note 1. Step 6 Select the desired units for energy rate and energy total. Step 7 Press Send to transmit calibration data to the transmitter. Step 8 Press Exit to close the Parameters Screen and return to the main screen. If required, press to view current programming of the transmitter. If required, press to enter factory default programming. Note: This will overwrite all settings. Step 5 Select the filter coefficients for flow and energy. See Note 3. Step 4 Select the type of temperature sensor. Step 3 Select the method of computing the temperature differential. See Note 2. Step 2 Select the desired unit for the temperature sensor. 10. After setting Model 340BN parameters, press Exit and re- turn to the Main screen. 11. Select Configure MS/TP to open the MS/TP Setup screen as shown below.