Programming
Example
If the meter was calibrated in US gallons, use PUL/GAL
Press ENTEO to save the K-factor unit and advance to the K FACTOR parameter
OTEE:
N
Unless otherwise specified, Blancett turbine flow meters are supplied
with K-factors measured in pulses per gallon (PUL/GAL) which will
automatically convert to your desired units of measure
Enter The Meter's K-Factor
OTEE:
N
The K-factor supplied with the meter or calculated from calibration data is
needed to set this parameter
To change the K-factor value, press OIGHT to select the digit that you wish to change Press
UP to increment the digit until it matches the meter’s K-factor Repeat this process until all
K-factor digits have been entered Press ENTEO to save the K-factor and advance to RATE/
TOTAL
Select the Units of Measure
The monitor is programmed with five common rate/total unit options The monitor shows
the rate/total unit that the display is currently set for If the current selection is correct, press
ENTEO once to advance to the DISPLAY FUNCTION parameter To change the unit, press UP
or OIGHT to scroll to the correct rate unit Press ENTEO to save and advance to the DISPLAY
FUNCTION parameter
Selection
Oate
Total
GPM/GAL
Gallons per minute
Gallons
LPM/LIT
Liters per minute
Liters
M3PH/M3
Cubic meters per hour
Cubic meters
M3PD/M3
Cubic meters per day
Cubic meters
BPD/BBL
Oil barrels per day
Oil barrels
Select the Display Function
The monitor displays RATE, TOTAL, alternates between BOTH rate and total or TEST If the
current selection is correct, press ENTEO to advance to the next parameter To change to an
alternate display mode, press UP or OIGHT to scroll to the correct display mode and press
ENTEO to save and advance to the TOTALIZER PULSE OUTPUT parameter
The TEST function acts like a frequency counter and displays the raw input frequency being
supplied to the frequency input terminals This is very useful when troubleshooting flow
problems
Page 7