Bartington PSU1 Operation Manual - page 17
BARTINGTON INSTRUMENT
Page 15 of 16
OM2443 ISSUE 1.0
Troubleshooting (contd.)
Fault
Possible Cause
Solution
With a sensor connected
all output signals are
faulty
Coupled incorrectly
Check position of AC/DC coupling switch
Balance mode set
incorrectly
Check position of balanced/unbalanced mode
switch
Defective sensor cable
Check sensor cable using relevant datasheet.
Return to BI if found to be defective
Sensor defective
Return PSU1 and sensor to Bartington
Instruments for repair
PSU1 defective
Return PSU1 and sensor to Bartington
Instruments for repair
With a sensor connected
one, or two output signals
are faulty
Defective sensor cable
Check sensor cable using relevant datasheet.
Return to BI if found to be defective
Sensor defective
Return PSU1 and sensor to Bartington
Instruments for repair
PSU1 defective
Return PSU1 and sensor to Bartington
Instruments for repair
If the solutions suggested above do not fix the fault, please contact the Bartington Instruments helpdesk.