Biosystems MultiPro Reference Manual - page 28
27
counter to advance past the maximum
possible alarm level for the specific channel,
and continue back to the lowest possible
setting.
To restore the default alarms
click the MODE button three
times in rapid succession
with “Default Alarms”
displayed. “Def Set” will be
shown to indicate that the
default alarm settings have been restored.
Once the alarm values have been set
correctly, click the MODE button three times
rapidly to return to the Main Menu.
6.6
Set cal gas
Once the set cal gas submenu has been
reached, the MultiPro will scroll through the
various calibration gas values screens and
the set default cal gas option.
Click the MODE button once
to advance through the
screens.
Hold the MODE button down
to advance the counter for a
specific calibration gas.
To set default calibration gas levels click the
MODE button three times in rapid succession
with “Default Cal Gas” displayed.
Once the calibration gas values have been
set correctly, click the MODE button three
times in rapid succession to return to the
main menu.
6.7
Set cal due
Once the set cal due
submenu is reached, the cal
due screen will be shown.
Hold the MODE button down to advance the
counter. The number of days shown is the
number of days that must pass following a
successful calibration before the calibration
due warning will be shown. The maximum
length of time between calibrations is 180
days.
To disable the calibration due warning, set
the interval to 0 days (OFF).
Once the calibration due interval has been
set correctly, click the MODE button three
times rapidly to return to the Main Menu.
Note: Biosystems recommends regular
verification of accuracy with calibration
gas. See Appendix B for details.
6.8
Set bump test due
Once the bump due submenu is reached, the
bump due screen will be shown. The
number of days shown is the number of days
that must pass following a successful
calibration before the bump due warning will
be shown.
Hold the MODE button down to advance the
counter in days. OFF will be followed by
DAILY, which will then be followed by 2 days
and so on. Once the interval passes 7 days,
it will proceed to 14 days and then to 21 days
and then to 30 days before it shows OFF
again. The maximum length of time between
scheduled bump tests is 30 days.
To disable the bump due warning, set the
interval to OFF.
Once the bump due interval
has been set correctly, click the
MODE button three times in
rapid succession to return to
the Main Menu.
6.9 Setup
accept
Once the settings in any of the submenus
have been changed as needed, return to the
Main Menu by clicking the MODE button
three times in rapid successtion. Once in the
Main Menu hold the MODE
button down for three seconds
to exit. The MultiPro will
provide a countdown to accept
changes to the settings.
Click MODE to accept the
changes and the instrument will
show “setup stored” and
proceed to turn itself off. Allow
the countdown to run out if you
do not wish to save the
changes.