Biosystems MultiPro Reference Manual - Certification Information
4
Certification Information
The MultiPro carries the following certifications:
UL Class I Division 1 Groups A,B,C,D Temp Code T4
CSA Class I, Division 1, Groups A,B,C,D Temp Code T4
(Per CSA C22.2 No. 152, only the combustible gas detection portion of this
instrument has been assessed for performance)
ATEX Certification:
II 2 G EEx ia d IIC T4
UL International DEMKO A/S 05 ATEX 0424116X
IECEx Certification:
Ex ia d IIC T4
IECEx UL 07.0003X
Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in
the MultiPro Reference Manual.
indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
indicates a potentially hazardous situation, which if not
avoided, may result in moderate or minor injury.
CAUTION used without the safety alert symbol indicates a potentially
hazardous situation which, if not avoided, may result in property damage.
Warnings and Cautions
1.
The MultiPro personal, portable gas detector has been
designed for the detection of dangerous atmospheric conditions. An alarm
condition indicates the presence of a potentially life-threatening hazard and should
be taken very seriously.
2.
In the event of an alarm condition it is important to follow
established procedures. The safest course of action is to immediately leave the
affected area, and to return only after further testing determines that the area is
once again safe for entry. Failure to immediately leave the area may result in
serious injury or death.
3.
The MultiPro
must be located in a non-hazardous location
whenever alkaline batteries are removed from the alkaline battery pack.
Removal of the alkaline batteries from the battery pack in a hazardous area may
impair intrinsic safety.
4.
Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer
E91-LR6, Eveready EN91 batteries in the alkaline battery pack. Substitution of
batteries may impair intrinsic safety.
5
.
To reduce the risk of explosion, do not mix old or used
batteries with new batteries and do not mix batteries from different manufacturers.
6.
Do not charge the MultiPro with any charger other than the
appropriate Biosystems MultiPro charger. Standard versions of the MultiPro must