Game Ready 500055 User Manual - page 19
Under Pressure:
The Control Unit
can’t reach its target maximum
compression. This often indicates
that there is a leak in the
pneumatic compression circuit,
either in the Connector Hose or
the Wrap itself. This is a Soft Error.
There is no danger to the person
being treated or the Control Unit,
and therapy will continue.
Once treatment has finished, try the
same pressure with a different Wrap,
if one is available. When you have
determined if the problem is isolated
to one Wrap or not, call Game Ready
customer service at 1.888.GameReady
(426.3732). You can continue to use your
system without harming the Control
Unit. You can even continue using the
affected Wrap, but this error will persist.
Low Flow Warning:
The Control
Unit has detected a temporary
interruption of fluid flow, but it
was able to correct the situation
by reducing the compression
setting. This is a Soft Error. There
is no danger to the person being
treated or the Control Unit, and
therapy will continue.
This error can often be corrected by
reapplying the Wrap so that it is snug
against the body part being treated.
Please consult your Wrap Use Guide for
more detailed instruction. If the problem
persists, call customer service at
1.888.GameReady (426.3732).
Over Pressure Alarm:
The Control
Unit has detected excessive
pressure in the air chamber.
In order to prevent any possible
harm to the person being treated,
the unit will stop therapy. This is
a Hard Error. The unit needs to be
turned off and restarted.
• Turn the Control Unit off and back on.
• Apply the same Wrap more tightly.
• Apply a different Wrap.
If the problem persists, call customer
service at 1.888.GameReady (426.3732).
Pressure Too High Alarm:
The
Control Unit has detected that the
Wrap has not properly deflated.
In order to prevent any possible
harm to the person being treated,
the unit will stop therapy. This is
a Hard Error. The unit needs to be
turned off and restarted.
This error often comes about when the
person being treated moves around dur-
ing treatment and puts pressure on the
air blader. If this is not the case try the
following:
• Turn the Control Unit off and back on.
• Apply the same Wrap more tightly.
• Apply a different Wrap.
If the problem persists, call customer
service at 1.888.GameReady (426.3732).
Low Flow Alarm:
The Control
Unit has detected an interruption
of fluid flow that it could not
correct. In order to prevent
possible damage to the fluid
pump, the unit will stop therapy.
This is a Hard Error. The unit
needs to be turned off and
restarted.
• Check the hose and Wrap for kinks that
might cut off fluid flow.
• Make sure there is water in the ice box.
• Make sure that the ice box filter is
clean.
• Reapply the Wrap so that it is snug
against the body part being treated.
If the problem persists, call customer
service at 1.888.GameReady (426.3732).
ERROR SCREEN
WHAT DOES IT MEAN? WHAT SHOULD I DO?
18
T R O U B L E S H O O T I N G
Soft Error:
When a Soft Error occurs, the system will continue to run, and will not fully
shut down. Therapy can continue even though the optimal levels are not being reached.
Hard Error:
When a Hard Error occurs, the system will shut itself down and will require
attention to start therapy again.
Soft Error
Soft Error
Hard Error
Hard Error
Hard Error