T R O U B L E S H O O T I N G
System will not turn on:
Check that the AC adapter is securely plugged into a working electrical outlet.
Control Unit turns off during use:
If the system current draw rises above 1.99 amps during operation, the Control Unit will
shut off to prevent damage to the pumps. Usually this is an indicator of a blockage in the
fluid or air flow. Check the Wrap and the Connector Hose to be sure there are no kinks,
creases or folds. Unplug the Control Unit to reset the system, and try the Wrap again. If
the Control Unit shuts off again, there is either a problem with the Wrap or the Control
Unit. Try a different Wrap to determine if it is the Control Unit or the Wrap and call Game
Ready customer service at 1.888.GameReady for further assistance.
Control unit leaks water when placed on its side:
1. If the Control Unit is on its side, check to be sure that the control panel is not upside
down. The unit is not designed to rest this way, and doing so will cause water to leak
through its air vent. While a minimal amount of water leakage should not damage the
system, immediately place the unit in its proper position: upright or on its side with the
rubber feet facing down.
2. If there is excess water on the rubber gasket around the mouth of the ice box, the lid
will not seal properly, and some of the water is likely to drip out when placed on its
side. Simply wipe off excess water from the gasket and close the lid securely.
3 Check the ice box gasket to be sure it’s seated properly around the lip of the ice box.
If it is not flat in the groove, push it down so that it is flush with the lip of the ice box.
If the gasket is damaged, call Game Ready customer service at 1.888.GameReady for
further assistance.