Good Energies B-CURE LLLT-808 Operating Instructions Manual - page 11
Indication
Status
B-CURE
®
LASER will automatically power off after 10 beeps because of flat batteries..
The “LOW” sign appears on the screen when laser power falls below the required level. Stop the
treatment. This situation can happen for the following reasons:
1. Battery power is too low.
2. The laser diode has reached the end of its life time and has to be replaced at the service center.
3. The device is being operated under environmental conditions not conforming to the specifications;
for example, the device is being operated at an excessive external temperature higher than 40ºC.
Recharge the batteries, turn the device on, and then see if the “LOW” sign still appears. If it does,
check if the environment is not too warm, and draw away from the warm area. In case the area
is not above 40ºC and the “LOW” sign is still on, the device is to be sent to the service center for
replacement of the laser diode.